摘要
目的:建立直接进样气相色谱法测定微量乙酸乙酯含量的分析方法。方法:通过对仪器参数进行优化,包括进样口温度、初始柱温、柱流速、分流比,确定最佳分析条件:色谱柱SH-PolarWax(30 m×0.32 mm 0.25μm),程序升温:初始温度为45℃,保持4min,以10℃/min升温至90℃,再以20℃/min升温至200℃;进样口温度220℃,分流比10∶1,流速1.00 m L/min。结果:在优化的色谱条件下,微量乙酸乙酯测试方法在0.01~1.00μg/mL范围内线性良好,相关系数(r)大于0.999,回收率在90%~110%范围内,RSD(n=3)均小于5%,检出限为0.004μg/mL,定量限为0.012μg/mL。结论:该方法具有检测时间短、灵敏度高的特点,研究结果可为乙酸乙酯含量检测方法研究提供数据支撑和参考。
Objective:To establish an analytical method for the determination of trace amounts of ethyl acetate by direct injection gas chromatography.Methods:The optimal analytical conditions were determined by optimizing the instrument parameters,including inlet temperature,initial column temperature,column flow rate,and split ratio:column SH-PolarWax(30 m×0.32 mm 0.25μm),programmed ramp-up:initial temperature of 45℃,hold for 4min,ramp-up to 90℃at 10℃/min,then ramp-up to 200℃at 20℃/min;inlet temperature of 220℃,flow rate of 1.00 mL/min,shunt ratio of 10∶1.Results:Under the optimized chromatographic conditions,the test method for the determination of trace amounts of ethyl acetate showed good linearity in the range of 0.01~1.00μg/mL with correlation coefficients(r)greater than 0.999,recoveries in the range of 90%~110%,RSDs(n=3)all less than 5%,detection limits of 0.004μg/mL and quantification limits of 0.012μg/mL.Conclusions:The method has the characteristics of short detection time and high sensitivity,and the research results can provide data support and reference for the research of ethyl acetate content detection method.
作者
樊美霞
汪毅
戴鹏
陈书蓉
占真雄
Fan Meixia;Wang Yi;Dai Peng;Chen Shurong;Zhan Zhenxiong(Jiangxi Sanxin Medtec Co.,Ltd.,Nanchang 330200,China)
出处
《广东化工》
CAS
2024年第8期104-108,共5页
Guangdong Chemical Industry
基金
南昌市重大科技攻关(洪科字[2019]243号)。
关键词
乙酸乙酯
气相色谱
条件优化
直接进样
外标法
ethyl acetate
gas chromatography
condition optimization
direct injection
external standard method