期刊文献+

基于电沉积富集-LIBS的水中Cd元素高灵敏稳定测量研究

Highly sensitive and stable measurement of Cd in water based on electrodeposition enrichment-LIBS technology
下载PDF
导出
摘要 针对水体痕量Cd元素的高灵敏稳定检测,基于Al片电沉积富集(ED)结合激光诱导击穿光谱(LIBS)检测的ED-LIBS水中Cd分析方法,实验研究了富集时间、富集电压、导电溶质KCl浓度以及电极表面砂纸打磨对Cd元素富集结果的影响。结果表明,经砂纸打磨Al片电沉积富集后,Cd元素富集稳定性得到提升,同时优化确定的富集电压、富集时间和KCl浓度分别为−2.4 V、1200 s和2 g/L。利用该方法,Cd元素检测限和相对标准偏差分别为0.0021 mg/L和3.30%,满足《生活饮用水卫生标准》指标检测需求,为水体中痕量Cd元素的高灵敏稳定分析提供了有力的检测手段。 Aiming at the highly sensitive and stable detection of trace amounts of Cd2+in water,an experimental study based on Al sheet electrodeposition enrichment(ED)combined with laser-induced breakdown spectroscopy(LIBS)to detect Cd in water analysis method(ED-LIBS)is carried out.The effects of enrichment time,enrichment voltage,concentration of conductive solute KCl,and sandpaper polishing on the electrode surface,on Cd element enrichment are analyzed.The result shows that the enrichment stability of Cd element is improved after the Al flakes for electrical-deposition being polished by sandpaper.The optimized enrichment voltage,enrichment time and concentration of KCl are−2.4 V,1200 s and 2 g/L,respectively.By using the proposed method,the detection limit and relative standard deviation of Cd element are 0.0021 mg/L and 3.30%,respectively,which meets the detection requirement of"Sanitary Standards for Drinking Water"and provides an optional method for the highly sensitive and stable analysis of trace Cd elements in water.
作者 疏阳 方丽 马明俊 孟德硕 赵南京 SHU Yang;FANG Li;MA Mingjun;MENG Deshuo;ZHAO Nanjing(Key Laboratory of Environmental Optics and Technology,Anhui Institute of Optics and Fine Mechanics,HFIPS,Chinese Academy of Sciences,Hefei 230031,China;University of Science and Technology of China,Hefei 230026,China;Key Laboratory of Optical Monitoring Technology for Environment,Anhui Province,Hefei 230031,China)
出处 《大气与环境光学学报》 CAS CSCD 2024年第3期314-321,共8页 Journal of Atmospheric and Environmental Optics
基金 国家自然科学基金(21735005,62001453) 安徽省重点研究和开发计划(201904a07020089)。
关键词 电沉积富集 激光诱导击穿光谱 水体重金属 高灵敏稳定测量 electrical-deposition enrichment laser-induced breakdown spectroscopy heavy metals in water highly sensitive and stable measurement
  • 相关文献

参考文献6

二级参考文献70

  • 1崔执凤,张先燚,姚关心,汪小丽,许新胜,郑贤锋,凤尔银,季学韩.铅黄铜合金激光诱导击穿谱特性的实验研究[J].原子与分子物理学报,2007,24(1):25-30. 被引量:26
  • 2李蕊,周琦.重金属污染与检测方法探讨[J].广东化工,2007,34(3):78-80. 被引量:19
  • 3许洪光,管士成,傅院霞,张先燚,许新胜,季学韩,凤尔银,郑荣儿,崔执凤.土壤中微量重金属元素Pb的激光诱导击穿谱[J].中国激光,2007,34(4):577-581. 被引量:48
  • 4K S Yun, J Gil, J Kim, et al. A miniaturized low-power wireless remote environmental monitoring system based on electrochemical analysis[J]. Sens.Actuators B, 2004,102( 1 ) : 27-34.
  • 5M Rawat, M C Z Moturi, V Subramanian. Inventory compilation and distribution of heavy metals in wastewater from small-scale industrial areas of Delhi, India[J]. J Environ Monit, 2003, 5(6):906-912.
  • 6J V Rios-Arana, E J Walsh, J L Gardea-Torresdey. Assessment of arsenic and heavy metal concentrations in water and sediments of the Rio Grande at E1 Paso-Juarez metroplex region[J]. Environment International, 2004,29 ( 7 ) : 957-971.
  • 7J Riondato, F Vanhaecke, L Moens, et al. Fast and reliable determination of (ultra-)trace and/or spectrally interfered elements in water by sector field ICP-MS[J]. J Anal At Spectrom, 2000, 15:341-345.
  • 8G L Calzoni, F Antognoni, E Pati, et al. Active biomonitoting of heavy metal pollution using Rosa rugosa plants[J]. Environ Pollut, 2007, 149: 239-245.
  • 9A M Zou, M L Chen, Y Shu, et al. Biological cell-sorption for separation/preconcentration of ultra-trace cadmium in a sequential injection system with detection by electrothermal atomic absorption spectrometry[J]. J Anal At Spectrom, 2007, 22(4): 392-398.
  • 10M H Devier, S Augagneur, H Budzinski, et al. One-year monitoring survey of organic compounds(PAHs, PCBs, TBT), heavy metals and biomarkers in blue mussels from the Arcachon Bay, France[J]. J Environ Monit,2005,7:224-240.

共引文献62