期刊文献+

MRI深度学习重建对乳腺脂肪抑制序列扫描时间和图像质量的影响

Impact of deep learning reconstruction in MRI on scan time and image quality of breast fat suppression sequence
下载PDF
导出
摘要 目的探讨磁共振成像(MRI)深度学习重建(DLR)对乳腺T_(2)加权(T_(2)WI)脂肪抑制序列扫描时间及图像质量的影响。方法招募30名女性志愿者为研究对象,采用平均采集次数(NAQ)为1和2的T_(2)WI脂肪抑制序列行乳腺MRI扫描,扫描时间分别为217 s和421 s,扫描完成获得Routine NAQ1和Routine NAQ2图像,Routine NAQ1组图像行DLR重建获DLR NAQ1图像,分析比较3组图像的信噪比(SNR)、对比噪声比(CNR)及临床医生主观定性评价资料。结果女性志愿者DLR NAQ1图像的SNR及CNR优于Routine NAQ1和Routine NAQ2图像(P<0.001),图像整体质量评分均优于Routine NAQ1和Routine NAQ2图像(P<0.001)。结论MRI DLR可缩短乳腺T2WI脂肪抑制序列的扫描时间,提高图像质量。 Objective To investigate the impact of deep learning reconstruction(DLR)in magnetic resonance imaging(MRI)on scan time and image quality of breast T_(2)-weighted(T_(2)WI)fat suppression sequence.Methods Thirty female volunteers were recruited as research subjects.Breast MRI scan was performed using T_(2)WI fat suppression sequence with an average number of acquisition(NAQ)of 1 and 2 as well as scan time of 217 s and 421 s,respectively.After scanning,Routine NAQ1 and Routine NAQ2 images were obtained.DLR was run on Routine NAQ1 image group to obtain DLR NAQ1 images.Signal-to-noise ratio(SNR)and contrast-to-noise ratio(CNR)of the images as well as the subjective and qualitative evaluation data from clinicians were analyzed and compared among three groups.Results SNR and CNR of DLR NAQ1 images of female volunteers were better than those of Routine NAQ1 and Routine NAQ2 images(P<0.001),and the overall quality score of the images is better than that of both Route NAQ1 and Route NAQ2 images(P<0.001).Conclusion DLR in MRI can shorten scan time of breast T 2WI fat suppression sequence and improve image quality.
作者 黄碧云 马卓雅 陈振涛 丁佳 冉孟新 刘世琛 李仕广 段庆红 HUANG Biyun;MA Zhuoya;CHEN Zhentao;DING Jia;RAN Mengxin;LIU Shichen;LI Shiguang;DUAN Qinghong(School of Medical Imaging,Guizhou Medical University,Guiyang 561113,Guizhou,China;Department of Image,Canon Medical SystemsCo.,Ltd of China,Beijing 100020,Beijing,China;Department of Imaging,Guizhou Medical University Affiliated Cancer Hospital,Guiyang 550000,Guizhou,China;Department of Imaging,the Second People's Hospital of Guiyang,Guiyang 550081,Guizhou,China)
出处 《贵州医科大学学报》 CAS 2024年第6期901-905,911,共6页 Journal of Guizhou Medical University
基金 贵州省科技计划项目(黔科合基础-ZK[2023]一般006,黔科合支撑[2021]一般451)。
关键词 磁共振成像 深度学习重建 脂肪抑制 平均采集次数 乳腺 扫描效率 magnetic resonance imaging deep learning reconstruction at suppression average number of acquisitions breast scan efficiency
  • 相关文献

参考文献9

二级参考文献57

  • 1陈万青,李霓,曹毛毛,任建松,石菊芳,陈宏达,李江,林春青,于欣阳,崔宏,代敏,赫捷.2013—2017年中国城市癌症早诊早治项目基线结果分析[J].中国肿瘤,2020,0(1):1-6. 被引量:110
  • 2刘小娟,翟仁友,王丽,蒋涛.磁共振成像在乳腺癌诊断中的应用[J].实用放射学杂志,2006,22(4):483-486. 被引量:7
  • 3曾庆师,李传福,孟祥水,崔谊,张训河.脂肪抑制FSPGR序列:胰腺病变定量和定性分析的ROC曲线评价[J].实用放射学杂志,2006,22(7):825-828. 被引量:2
  • 4Boutry S, Burtea C, Laurent S, et al. Magnetic resonance imaging of inflammation with a specific seleetin-targeted contrast agent. Magn Reson Med, 2005,53 (4) :800-807.
  • 5Li XB, Du XK, Huo TL, et al. Specific targeting of breast tumor by octreotideoctreotide-conjugated ultrasmall superparamagnetieiron oxide particles using a clinical 3.0-Tesla magnetic resonance scanner. Acta Radiol, 2009,50(6) :583-594.
  • 6Wang LY, Park HY, Lira SII, et al. Core@shell nanomaterials: gold-coated magnetic oxide nanoparticles. 3 Mater Chem, 2008,18(23):2629-2635.
  • 7Jabbari E. Targeted delivery with peptidomimetic conjugated selfassembled nanoparticles. Pbarm Res, 2009,26(3):612-630.
  • 8Artemov D, Mori N, Okollie B, et al. MR molecular imaging of the Her-2/neu receptor in breast cancer cells using targeted iron oxide nanoparticles. Magn Reson Med, 2003,49(3):403-408.
  • 9Yang LL, Peng XH, Wang YA, et al. Receptor-targeted nanoparticles for in vivo imaging of breast cancer. Clin Cancer Res, 2009,15(14) :4722-4732.
  • 10Oude Munnink TH, NagengaSt WB, Brouwers AH, et al. Molecular imaging of breast cancer. Breast, 2009,18(Suppl 3):S66- S73.

共引文献124

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部