期刊文献+

心房颤动时损伤电流指导心房电极植入1例

下载PDF
导出
摘要 患者女,72岁,诊断为“病态窦房结综合征、阵发性心房颤动”等,反复发作头昏、视物黑矇,考虑由心动过缓所致,拟行起搏器治疗,但术前患者心房颤动发作,术前、术中均未转为窦性心律。为了避免右心室起搏的危害,术中行左束支区域起搏,并在损伤电流指导下植入心房电极,术后转为窦性心律,心房电极起搏功能正常,起搏参数良好。术后1个月复诊,测试心房、心室电极参数良好。
出处 《实用心电学杂志》 2024年第4期362-364,共3页 Journal of Practical Electrocardiology
基金 四川省卫健委干保课题(川干研2023-201)。
  • 相关文献

参考文献5

二级参考文献29

  • 1徐伟,吉文庆,李晓宏,黄从新.主动固定心房导线的起搏阈值在植入时和随访期的改变[J].中华心律失常学杂志,2007,11(2):124-125. 被引量:6
  • 2Healey JS,Toff WD,Lamas GA,et al.Cardiovascular outcomes with atrial-based pacing compared with ventricular pacing:meta-analysis of randomized trials,using individual patient data.Circulation,2006,114:11-17.
  • 3Fyfe T,Maclntyre PD,Robinson JF,et al.Blind atrial pacing for patients with sinus node disease who develop atrial fibrillation during permanent pacemaker implantation.Int J Cardiol,1997,58:188-191.
  • 4Kindermann M,Frohlig G,Berg M,et al.Atrial lead implantation during atrial flutter or fibrillation? Pacing Clin Elactrophysiol,1998,21:1531-1538.
  • 5Wood MA,Moskovljevic P,Stambler BS,et al.Comparison of bipolar atrial elactrngram amplitude in sinus rhythm,atrial fibrillation,and atrial flutter.Pacing Clin Electrophysiol,1996,19:150-156.
  • 6Lewalter T,Schimpf R,Kulik D,et al.Comparison of spontaneous atrial fibrillation electrogram potentials with the P-wave electrogram amplitude in dual chamber pacing with unipolar atrial sensing.Europace,2000,2:136-140.
  • 7Palma EC,Kedarnath V,Vankawalla V,et al.Effect of varying atrial sensitivy,AV interval,and detection algorithm on automatic mode switching.Pacing Clin Electrophysiol,1996,19:1734-1739.
  • 8Rieci R,Puglisi A,Azzolini P,et al.Reliability of a new algorithm for automtic mode switching from DDDB to DDIR pacing mode in sinus node disease patients with chronotropic incompetence and recurrent paroxysmal atrial fibrillation.Pacing Clin Electrophysiol,1996,19:1719-1723.
  • 9Neuzner J,Sperzel J,Pitsehner HF,et al.Bipolar atrial sensing thresholds in sinus rhythm and atrial tachyarrhythm ias.A comparative analysis in patients with DDDR pacemakers.Europace,1999,1:135-139.
  • 10Wolfhard UF,Eichstaedt HC,Sack S,et al.Atrial fibrillatory electrogram measurement allows atrial lead placement in patients who develop atrial fibrillation during permanent dual chamber pacemaker implantation.Pacing Clin Electrophysiol,1998,21:2300-2303.

共引文献18