摘要 在利用镍触媒甲烷化气相色谱法分析空气中一氧化碳的实验中,随着高分子微球色谱柱使用周期和频次的增加,一氧化碳测定结果呈现逐渐增高的现象,特别是对含有微颗粒物受限空间的气体样品分析,一氧化碳结果严重异常。通过实验对比不同气体溶剂本底的样品,对比不同使用频次的色谱柱,找到其主要干扰因素是微颗粒物改变柱效形成色谱峰干扰。 In the test of carbon monoxide analysis in air by nickel-catalyzed methanation gas chromatography,the results of carbon monoxide determination gradually increased with the increase of the use period and frequency of polymer microsphere column,especially for the analysis of gas samples containing small particles in confined space,the results of carbon monoxide were seriously abnormal.By comparing the samples with dfferent gas solvent background and chromatographic columns with dfferent use frequency,it is found that the main interference factor is the chromatographic peak interference caused by the change of column effect of fine particles.
作者 林志 甘玉明 李志新 LIN Zhi;GAN Yuming;LI Zhixin(Quality Inspection Center,PetroChina Jilin Petrochemical Company,Jilin,Jilin 132011 China)
出处 《石油工业技术监督》 2024年第8期40-42,共3页 Technology Supervision in Petroleum Industry
关键词 一氧化碳 氧气 高分子微球 微颗粒物 carbon monoxide oxygen polymer microspheres fine particulate matter
  • 相关文献

参考文献10

二级参考文献37

  • 1刘锋,李思慧.一氧化碳中微量CO_2的热导色谱分析[J].天然气化工—C1化学与化工,1996,21(2):51-53. 被引量:3
  • 2GB/T 5274-2008 气体分析校准用混合气体的制备称量法[S].
  • 3顾惠样,阎宝石.气相色谱实用手册(第2版)[M].北京:化学工业出版社.1990:99.
  • 4A. T. Blades. J. Chxomatogr. Sci. , 1976,14( 1 ) :45.
  • 5Armond. T. American Laboratory, may, 1976,32:625.
  • 6Horst Hachenberg, Industrical Gas chromatogT. Trace Aanlysis, P. 68. Heyoer & Son Ltd. ,1973.
  • 7P. R. Wentrek, B. J. Wood, H. Wise. J. Catal. , 1976,43:363.
  • 8GB/T7445-1995.纯氢、高纯氢、超纯氢[S].[S].,..
  • 9GB/T 14599-93,高纯氧[S].
  • 10GB/T8980-1996高纯氮[S].

共引文献40

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部