UPLC-MS/MS法及均相酶免疫法测定人全血西罗莫司浓度

Determination of sirolimus concentration in human whole blood by UPLC-MS/MS and homogeneous enzyme immunoassay
下载PDF
导出
摘要 目的建立高效液相色谱-串联质谱(UPLC-MS/MS)法监测人全血西罗莫司浓度,并将检测结果与均相酶免疫法进行比较。方法100μL全血样品加入10μL西罗莫司-D3内标、100μL 0.1 mol·L^(-1)ZnSO4溶液、300μL甲醇,涡旋离心后,上清液5μL进样。以0.1%甲酸-甲醇为有机相,0.1%甲酸-水(含10 mmol·L^(-1)乙酸铵)为水相,流速0.5 mL·min^(-1),FastCore Super C18(2.1 mm×100 mm,2.6μm)色谱柱进行梯度洗脱分离。在多反应监测(MRM)模式下采用电喷雾离子源(ESI)正离子化方式检测,西罗莫司、内标的定量离子对分别为m/z 931.6>864.6,m/z 934.6>864.6。进行方法学考察(包括特异性、定量下限、标准曲线、准确度、精密度、残留效应、基质效应、稳定性和回收率)。测定94例器官移植患者的西罗莫司浓度,与均相酶免疫法结果进行比较。结果西罗莫司标准曲线浓度范围为2.5~50 ng·mL^(-1)。西罗莫司的日内、日间准确度相对误差(RE)在-5.3%~7.5%,变异系数(CV)小于11.8%。低、中、高不同浓度的质控回收率和基质效应一致,经内标校正后回收率为100.22%~108.56%,经内标归一化后基质效应为95.45%~96.48%,CV小于7.0%。全血样本室温放置3 d,4℃放置3 d,-20℃放置20 d,反复冻融3次,处理后样本自动进样器放置20 h,室温放置24 h结果稳定。Passing-Bablok回归分析及Bland-Altman结果表明均相酶免疫法与UPLC-MS/MS两种西罗莫司测量方法存在比例偏差。结论本研究建立了一种简单、高效的UPLC-MS/MS方法检测人全血西罗莫司浓度,均相酶免疫法与UPLC-MS/MS法检测人全血西罗莫司存在比例偏差。UPLC-MS/MS方法更加准确,是免疫抑制剂治疗药物监测检测金标准。 Objective To quantify sirolimus concentration in human whole blood by UPLC-MS/MS and to evaluate the agreement with the homogeneous enzyme immunoassay.Methods Human whole blood(100μL)was spiked with 10μL sirolimus-D3(the internal standard,IS),100μL 0.1 mol·L^(-1)ZnSO4 and 300μL methanol.After the vortex and centrifugation(14000 r·min^(-1)),5μL supernatant was injected.Water(containing 10 mmol·L^(-1)ammonium acetate and 0.1%formic acid,solvent A)and methanol(containing 0.1%formic acid,solvent B)were used as the eluent,with the gradient elution at 0.5 mL·min^(-1)in FastCore Super C18 column(2.1 mm×100 mm,2.6μm).The multiple reaction monitoring(MRM)with electrospray ionization(ESI)in the positive-ion mode was used.m/z 931.6>864.6 and m/z 934.6>864.6 were used for sirolimus and the IS quantitation.Method validation,including selectivity,LLOQ,calibration curve,accuracy,precision,remaining-effect,matrix effect,stability and recovery was evaluated.Passing-Bablok regression analysis and Bland-Altman plot were used to assess the agreement between UPLC-MS/MS and homogeneous enzyme immunoassay.Results The calibration ranged 2.5~50 ng·mL^(-1)for sirolimus.The relative error of intra-and inter-day accuracy of sirolimus ranges from-5.32%to 7.46%,and the coefficient of variation was less than 11.8%.The IS normalized recovery and matrix effect were 100.22%~108.56%and 95.45%~96.48%with coefficient of variation less than 7.0%.Sirolimus was stable in the test conditions.Comparison between UPLC-MS/MS with homogeneous enzyme immunoassay showed a deviation in proportion.Conclusion UPLC-MS/MS method for the determination of sirolimus in human whole blood is rapid,easy and reliable.A proportion deviation in the two methods is observed.
作者 杨华 王兰 宋微微 杨晓娟 冀召帅 艾超 YANG Hua;WANG Lan;SONG Wei-wei;YANG Xiao-juan;JI Zhao-shuai;AI Chao(Department of Pharmacy,Beijing Tsinghua Changgung Hospital,School of Clinical Medicine,Tsinghua University,Beijing 102218)
出处 《中南药学》 CAS 2024年第8期2053-2057,共5页 Central South Pharmacy
关键词 西罗莫司 UPLC-MS/MS方法 方法学开发 方法学验证 sirolimus UPLC-MS/MS method development method validation
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部