Mobetron 2000型术中放疗设备的性能调试

Performance debugging of the Mobetron 2000 intraoperative radiotherapy device
下载PDF
导出
摘要 目的探索新型术中放疗设备现场检测方法,为该类设备检测标准研制提供实验支撑数据。方法以某医院1台Mobetron 2000型术中移动式加速器为研究对象,参照标准,使用MP3-XS小型三维水箱和Dose2剂量仪等检测设备,检测电子线照射野均整度、对称性、最大吸收剂量比等指标。结果该术中移动式加速器均整度为2.5~3.7 mm,对称性为100.4%~101.1%,最大吸收剂量比为102.8%~105.3%,剂量偏差为-2.4%~-0.5%,重复性均为0.4%,线性为0.6%~1.0%,日稳定性为-0.4%~0.28%,6、9、12 MeV能量点杂散辐射分别为0.3%、0.4%、0.5%。结论在非标情况下,现有放疗检测设备测量术中移动式加速器结果可为标准研制提供参考。 Objective To investigate the approach for field testing of a novel intraoperative radiotherapy device,so as to provide experimental support data for development of device testing criteria.Methods A Mobetron 2000 intraoperative mobile accelerator in a hospital was selected,and the uniformity,symmetry and the maximum absorbed dose ratio of electron beam irradiation fields were determined with the MP3-XS Water Phantom System and Dose2 Dosimeter according to the national reference.Results The uniformity,symmetry,the maximum absorbed dose ratio,dose deviation,repeatability,linearity and daily stability of the intraoperative mobile accelerator were 2.5 to 3.7 mm,100.4%to 101.1%,102.8%to 105.3%,-2.4%to-0.5%,0.4%,0.6%to 1.0%and-0.4%to 0.28%,and the stray radiations of 6,9,12 MeV energy sites were 0.3%,0.4%and 0.5%.Conclusions The currently available radiotherapy testing results of intraoperative mobile accelerators may provide insights into formulation of criteria under non-standard circumstances.
作者 徐小三 杜翔 王进 XU Xiao-san;DU Xiang;WANG Jin
出处 《江苏预防医学》 CAS 2024年第3期355-357,369,共4页 Jiangsu Journal of Preventive Medicine
关键词 移动式加速器 术中放疗 三维水箱 剂量计 Mobile accelerator Intraoperative radiotherapy Three-dimensional water tank Dosimeter
  • 相关文献

参考文献5

二级参考文献26

  • 1王军良,周振山,盛洪国,齐伟华.IAEA TRS398与TRS277应用于加速器输出量校准的比较[J].中国医学物理学杂志,2011,28(3):2594-2596. 被引量:11
  • 2姜瑞瑶,黄国锋,熊霏,傅深.术中放疗深度剂量、表面剂量及漏射线的剂量研究[J].中国医学物理学杂志,2007,24(3):168-169. 被引量:5
  • 3张红志 见:胡逸民 主编.高能电子束射野剂量学[A].见:胡逸民,主编.肿瘤放射治疗学[C].北京:原子能出版社,1999.253-254.
  • 4Beddar AS, Kubu ML, Domanovic MA, et al. A new approach to intraoperative radiation therapy. AORN J,2001,74 :500-505.
  • 5Mills MD, Fajardo LC, Wilson DL, et al. Commissioning of a mobile electron accelerator for intraoperative radiotherapy. J Appl Clin Med Phys,2001,2:121-130.
  • 6Beddar AS. Stability of a mobile electron linear accelerator system for intraoperative radiation therapy. Med Phys,2005 ,32 :3128-3131.
  • 7Beddar AS, Domanovic MA, Ellis RJ, et al. Using intraoperative radiation therapy-a case study. AORN J, 2003,77:412-417.
  • 8Beddar AS, Biggs PJ, Chang S, et al. Intraoperative radiation therapy using mobile electron linear accelerators. Report of AAPM Radiation Therapy Committee Task Group No. 72. Med Phys,2006, 33 : 1476-1489.
  • 9Sawaki M, Sato S, Noda S, et al: Phase Ⅰ/Ⅱ study of intraoperative radiotherapy for early breast cancer in Japan. Breast Cancer, 2011.
  • 10Kimple RJ, Klauber-DeMore N, Kuzmiak CM, et al: Cosmetic outcomes for accelerated partial breast irradiation before surgical excision of early-stage breast cancer using single-dose intraoperative radiotherapy. Int J Radiat Oncol Biol Phys 79: 400-407, 2011.

共引文献32

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部