一种新型人工膝关节胫骨衬垫的体外磨损分析

In vitro wear analysis of a new type of tibial liner for artificial knee joint
下载PDF
导出
摘要 对添加十二烷基没食子酸酯的新型高交联聚乙烯材料(XLPE-DG)制造的膝关节胫骨衬垫进行体外磨损测试,研究其磨损质量损失(磨损量)、磨损率和磨损颗粒的特征(数量、大小和形状),分析其在不同磨损阶段的磨损性能,为国产医用聚乙烯材料的研发及临床应用提供数据支持。按照国际标准化组织(ISO)14243-1和14243-2标准方法,选取4件全膝关节置换假体为测试样品(3件试验组和1件对照组),试验中的润滑剂为小牛血清,利用膝关节模拟机对XLPE-DG衬垫进行循环磨损测试,分别计算100万次循环(1 Mc)、300万次循环(3 Mc)及500万次循环(5 Mc)磨损后的磨损量和磨损率,使用图像分析软件对扫描电镜获得的磨屑图像分析,分析磨屑的特征(数量、大小和形状)。通过5 Mc磨损测试后XLPE-DG衬垫的磨损量为(23.86±0.75)mg,磨损率为(4.39±0.75)mg/Mc,3~5 Mc阶段的磨损量与1~3 Mc阶段磨损量接近(8.79 mg vs 8.74 mg);磨屑颗粒数量随着磨损次数的增加而减少;5 Mc后XLPE-DG衬垫的磨屑颗粒等效圆面积直径(ECD)平均值为0.13μm(最小值0.04μm,最大值0.8μm),且(77.11±6.89)%的颗粒ECD平均值<0.2μm,ECD平均值随着磨损次数的增加而下降;5 Mc磨损后衬垫磨屑体态比(AR)平均值为2.8(最小1.0,最大9.9),其中77.9%的颗粒AR值<4,以圆形粒状颗粒为主。XLPE-DG衬垫表现出良好的耐磨损性能,在临床应用中能够减少发生假体周围炎症、骨溶解和假体松动的风险,具有良好的临床应用潜力。 In vitro wear tests were conducted on knee tibial liner manufactured from a new highly cross-linked polyethylene material(XLPE-DG)with dodecyl gallate.The characteristics(quantity,size and shape)of wear mass loss(wear volume),wear rate and wear particles were studied,and the wear performance in different wear stages was analyzed,so as to provide data support for the research and development and clinical application of domestic medical polyethylene materials.In accordance with the International Organization for Standardization(ISO)14243-1 and 14243-2 standard methods,4 total knee arthroplasty prostheses were selected as test samples(3 in the test group and 1 in the control group).The lubricant in the test was calf serum,and the XLPE-DG liner was tested for cyclic wear using a knee joint simulator.The wear volume and the wear rate after 1 million cycles(1 Mc),3 million cycles(3 Mc),and 5 million cycles(5 Mc)were calculated respectively.The abrasive debris images obtained by the scanning electron microscope(SEM)was analyzed by using image analysis software to analyze the characteristics of the debris of the wear debris(number,size and shape).After passing the 5 Mc wear test,the wear amount of XLPE-DG liner was(23.86±0.75)mg,with a corresponding wear rate of(4.39±0.75)mg/Mc and the wear amount during the 3–5 Mc stage was similar to that of the 1–3 Mc stage(8.79 mg vs.8.74 mg).The number of wear debris decreased with the increase of the number of wear times.The average value of the equivalent circular area diameter(ECD)of the wear debris of XLPE-DG liner after 5 Mc was 0.13μm(minimum value 0.04μm,maximum value 0.8μm),and the average ECD of 77.11%±6.89%particles was<0.2μm,the average value of ECD decreased with the increase of the number of wear times.The average value of the aspect ratio(AR)of the liner after 5 Mc wear was 2.8(minimum 1.0,maximum 9.9),of which 77.9%had an AR value of<4 and the particles are mainly round.The XLPE-DG liner demonstrated excellent wear resistance and can reduce the risk of periprosthetic inflammation,osteolysis,and prosthetic loosening in clinical applications,and have good clinical application potential.
作者 闫师通 王建军 苏健 Yan Shitong;Wang Jianjun;Su Jian(Department of Material No.2,Beijing Institute of Medical Device Testing,Beijing 101111,China)
出处 《中国医学装备》 2024年第9期171-177,共7页 China Medical Equipment
关键词 膝关节假体 体外磨损 磨损率 磨损量 磨屑分析 Knee prosthesis In vitro wear Wear rate Wear volume Wear debris analysis
  • 相关文献

参考文献4

二级参考文献24

  • 1张超远,周勇刚,张建党,宋应超,李振武.特制全膝关节置换术后感染的原因分析及防治[J].军医进修学院学报,2005,26(3):186-187. 被引量:2
  • 2吴海山.对全膝关节置换术现状的共识和研究方向[J].中华骨科杂志,2005,25(7):414-417. 被引量:21
  • 3关振鹏,吕厚山,陈彦章,宋奕宁,秦秀龙,姜军.影响人工关节置换术后下肢深静脉血栓形成的临床风险因素分析[J].中华外科杂志,2005,43(20):1317-1320. 被引量:133
  • 4樊新甫,杨凤云,董英海.人工关节置换术后的骨吸收和骨溶解[J].临床骨科杂志,2006,9(1):92-94. 被引量:5
  • 5Walker PS,Sathasivam S.The design of guide surfaces for fixed-bearing and mobile-bearing knee replacements.J Biomech.1999;32(1):27-34.
  • 6Stiehl JB,Dennis DA,Komistek RD,et al.In vivo kinematic comparison of posterior cruciate ligament retention or sacrifice with a mobile bearing total knee arthroplasty.Am J Knee Surg.2000;13(1):13-18.
  • 7Barrack RL,Rorabeck C,Partington P,et al.The results of retaining a well-fixed patellar component in revision total knee arthroplasty.J Arthroplasty.2000;15(4):413-417.
  • 8Geerts WH,Pineo GF,Heit JA,et al.Prevention of venous thromboembolism:the Seventh ACCP Conference on Antithrombotic and Thrombolytic Therapy.Chest.2004;126(3 Suppl):338S-400S.
  • 9Bergquist D.Postoperative hromboembolism:frequentecy,eiology,prophylaxis.Berlin:Springer Verlag.1983:129-145.
  • 10Rorabeck CH,Taylor JW.Periprosthetic fractures of the femur complicating total knee arthroplasty.Orthop Clin North Am.1999;30(2):265-277.

共引文献1385

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部