目的:建立电感耦合等离子体质谱法(ICP-MS)测定聚山梨酯80中24种元素含量,并进行分析评价。方法:采用ICP-MS法进行测定,通过SPSS 26软件进行统计分析,参考ICH Q3D进行元素杂质风险评估。结果:各元素在对应浓度范围内线性关系良好(r=0.997~0.999);回收率88.81%~134.64%(RSD<5%,n=6);样品中建议评估的元素均未超过推荐控制阈值。结论:该方法准确可靠、操作方便,品种整体元素杂质风险较低。
Objective:To establish an inductively coupled plasma-mass spectrometry(ICP-MS)method for the determination of 24 elements in polysorbate 80,and evaluate the results.Methods:The elements were determined by ICP-MS,statistical analysis was conducted using SPSS 26,and the risk assessment was carried out with reference to ICH Q3D guidelines.Results:The standard curves showed good linear relationship in the corresponding concentration range(r=0.997-0.999).The recoveries were 88.81%-134.64%(RSD<5%,n=6).None of the elements recommended for evaluation exceeded the control threshold.Conclusion:The method is accurate,reliable and easy-to-operate,and the risk of elemental impurities in this sample is low.