期刊文献+

两种常用品牌双能X线骨密度仪应用标准化中国库的检出率对比

Comparison of the detection rate of two commonly used brands of dual-energy X-ray bone densitometry in standardized Chinese database
下载PDF
导出
摘要 目的回顾分析某三甲医院两种常用品牌双能X线吸收测量仪(dual energy X-Ray absorptiometry,DXA)应用标准化中国库(依据近期研究成果,首次试用)前后的骨质疏松检出率差异,以验证该标准化数据库的使用效果。为实现不同品牌DXA T值标准化提供理论和数据支持。方法收集2021年6月11日—2022年10月20日GE-Prodigy以及2021年10月1日—2022年9月9日Hologic Horizon-W测量的门诊30岁以上中国患者的骨密度结果。分别获得腰椎(L1~4)和单侧髋应用“中国库”和“标准化中国库”的T值,并依据T值计算两种品牌DXA应用标准化数据库前后的检出率。结果两种品牌DXA应用标准化数据库前后的L1~4检出率分别相差136.99%和4.58%;单侧髋检出率分别相差58.69%和22.99%。结论GE-Prodigy与Hologic Horizon-W对骨质疏松的检出率差异较大,标准化中国库的应1用降低了这两种国内常用品牌DXA的检出率的差异,将很大程度上提高骨质疏松诊断的同质性。 Objective The difference in the osteoporosis detection rate before and after the application of the standardized Chinese database(based on recent research results,the first trial)of two commonly used brands of dual-energy X-ray absorptiometer(DXA)in a third class hospital was reviewed and analyzed,so as to verify the application effect of the standardized database and provide theoretical and data support for the standardization of T values of different brands of DXA.Methods BMD results from out-patient Chinese patients over 30 years of age measured by GE-Prodigy from June 11,2021 to October 20,2022 and Hologic Horizon-W from October 1,2021 to September 9,2022 were collected.The T-values of lumbar spine(L1~4)and unilateral hip were obtained,and the detection rates of DXA before and after the application of the standardized database were calculated.Results The difference of L1~4 detection rates between the two brands before and after applying the standardized database was 136.99%and 4.58%,respectively.The detection rate of single hip was 58.69%and 22.99%respectively.Conclusion There were significant differences in the positive rates of osteoporosis between GE-Prodigy and Hologic Horizon-W.The application of standardized databases reduced the differences in positive rates between these two commonly used types of DXA devices in China and greatly improved the homogeneity of osteoporosis diagnosis.
作者 轩艳姣 倪晓龙 孙鸣珂 程晓光 闫东 Xuan Yanjiao;Ni Xiaolong;Sun Mingke;Cheng Xiaoguang;Yan Dong(Department of Radiology,Beijing Jishuitan Hospital of Capital Medical University,Beijing 100035,China)
出处 《现代仪器与医疗》 CAS 2024年第5期35-39,共5页 Modern Instruments & Medical Treatment
基金 北京市医院管理中心临床医学发展专项经费资助(ZYLX202107)。
关键词 骨密度 双能X射线吸收法 标准化数据库 检出率 Bone mineral density Dual-energy X-ray absorptionmetry Standardized database Positive rate
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部