摘要 随着我国心律植入装置应用数量不断增加,出现导线感染、断裂等相关问题需要拔除电极导线的病例也日趋增多。电极导线与心肌组织或静脉系统粘连增加了电极导线拔除技术的难度。文章对锁定导丝、机械扩张鞘和圈套器等经静脉电极导线拔除的三种无源医疗器械进行概述,并对审评关注点及风险点进行浅析。 As the use of cardiac rhythm management devices continues to rise in China,the number of cases requiring lead extraction due to complications such as lead infection,lead fracture,and vascular occlusion is also increasing.The adhesion of the lead to myocardial tissue or the venous system complicates the lead extraction procedure.This article provides an overview of three passive medical devices used for transvenous lead extraction:the locking stylet,mechanical dilation sheath,and snare device.It also discusses key considerations and risks associated with their regulatory review.
作者 许耘 程茂波 XU Yun;CHENG Mao-bo(Center for Medical Device Evaluation.NMPA,Beijing 100076)
出处 《中国医疗器械信息》 2024年第19期18-20,共3页 China Medical Device Information
关键词 心律植入装置 电极导线拔除技术 锁定导丝 机械扩张鞘鞘 圈套器 cardiac rhythm management devices lead extraction techniques locking stylet mechanical dilation sheath snare device
  • 相关文献

参考文献4

二级参考文献18

  • 1Larry MB, Andrew EE, Christopher CE, et al. Update on cardiovascu- lar implantable electronic device infections and their management:a scientific statement from the American Heart Association [ J ]. Circu- lation,2010,121:458.
  • 2Margey R. Cardiac implantable electronic device infections:the ene- my that lurks beneath the skin[ J]. J Long Term Eft Med Implants, 2010,20 (3) :203.
  • 3Diemberger I, Biffi M, Martignani C, et al. From lead management to implanted patient management:indications to lead extraction in pace- maker and cardioverter- defibrillator systems [ J ]. Expert Rev Med Devices,2011,8(2) :235.
  • 4Moak JP, Mercader MA, Berul CI. Acute complications of electro- physiology and pacing procedures: identification and management [ J ]. Minerva Cardioangio1,2010,58 (4) :485.
  • 5Calvagna GM, Evola R, Valsecchi S. A complication of pacemaker lead extraction: pulmonary embolization of an electrode fragment [ J]. Europace,2010,12(5 ) :613.
  • 6郭继鸿,王斌主编.人工心脏起搏技术[M].第1版.北京:辽宁科学技术出版社,2008.104-115.
  • 7Farooqi FM, Talsania S, Hamid S, et al. Extraction of cardiac rhythm devices: indications, techniques and outcomes for the removal of pacemaker and defibrillator leads [ J ]. Int J Clin Pract, 2010,64 (8) :1 140.
  • 8Henrikson CA,Brinker JA. How to prevent, recognize, and manage complications of lead extraction. Part I : avoiding lead extraction-in- fectious issues [ J ]. Heart Rhythm,2008, 5 (7) : 1 083.
  • 9陈泗林,林纯莹,费洪文,刘烈,陈东骊,梁远红,吴书林.心脏再同步治疗起搏参数优化对血流动力学的影响及随访[J].中华心律失常学杂志,2009,13(2):133-136. 被引量:6
  • 10华伟,丁立刚,张澍,陈柯萍,王靖,王方正.远程家庭监测功能在心脏再同步治疗中的应用[J].中华心律失常学杂志,2010,14(3):169-171. 被引量:11