基于拟合光源的荧光摄像系统性能测试与评价方法

Performance Evaluation Methods Based on Spectral Fitting for Near-Infrared Fluorescence Imaging
下载PDF
导出
摘要 基于吲哚菁绿造影剂的荧光摄像系统在手术和研究中快速普及,但目前荧光性能的检测方法存在复现与溯源的问题,无法满足标准化评价荧光摄像系统的新需求。该研究提出基于拟合光源的荧光性能评价方法,定义三类荧光性能参数并通过对比验证。基于拟合光源的荧光性能评价方法为建立荧光图像质量评估标准提供可行的标准参考。 The use of near-infrared fluorescence imaging with indocyanine green in surgery and research is becoming increasingly widespread.Nowadays,the performance test methods for near-infrared fluorescence imaging have the problems of bad repeatability and untraceability.In order to address the emerging need for standardization of near-infrared fluorescence imaging systems,it is necessary to identify the best methods suitable for objective,quantitative testing of image quality characteristics.Based on fluorescence imaging literature,we have developed and demonstrated three test methods for evaluation of fluorescence image quality,which are based upon simulation of near-infrared fluorescence.These methods have the potential to facilitate other quantitative comparison of device performance,with the long-term goal of establishing standards for fluorescence image quality assessment.
作者 张沁园 徐张恒 颜青来 陈德宝 沈焕波 ZHANG Qin-yuan;XU Zhang-heng;YAN Qing-lai;CHEN De-bao;SHEN Huan-bo(Zhejiang Institute of Medical Device Testing,Zhejiang Hangzhou 310019)
出处 《中国医疗器械信息》 2024年第19期40-43,55,共5页 China Medical Device Information
关键词 荧光摄像 拟合荧光 性能检测 near-infrared fluorescence imaging simulation of near-infrared fluorescence performance test methods
  • 相关文献

参考文献3

二级参考文献14

  • 1Engel E, Schraml R, Maisch T, et al. Light-Induced decom- position of indocyanine green [ J ]. Invest Ophth Vis Sci, 2008,49(5) : 1777-1783.
  • 2Fernandez A, Manchanda R, Lei T, et al. Comparative study of the optical and heat generation properties of IR820 and indocyanine green [ J ]. Mol Imaging ,2012,11 (2) :99 -113.
  • 3Penha F M, Rodrigues E B, Maia M, et al. Biochemical a- nalysis and decomposition products of indocyanine green in relation to solvents, dye concentrations and laser exposure [J]. Ophthalmologica,2013,230(2) :59-67.
  • 4Ogata F, Azuma R, Kikuchi M, et al. Novel lymphography using indocyanine green dye for near-infrared fluorescencelabeling[ J]. Ann Plas Surg,2007,58 (6) :652-655.
  • 5Raabe A,Nakaji P, Beck J, et al. Prospective evaluation of surgical microscope - integrated intraoperative near-infra- red indoeyanine green videoangiography during aneurysm surgery [ J ]. J Neurosurg , 2005,103 (6) :982-989.
  • 6] Nomikou N, Sterrett C, Arthur C, et al. The effects of ultra- sound and light on indocyanine-green-treated tumour cells and tissues [ J ]. Chem Med Chem, 2012,7 ( 8 ) : 1465-1471.
  • 7刘晶静,曾超,冯素玲.荧光素染料与富马酸酮替芬的光谱研究及应用[J].化学研究与应用,2010,22(3):301-305. 被引量:7
  • 8刘彬,孔玉梅,王新,刘冰弥,王丹,王冬晶,王君.声动力疗法和声敏剂研究进展[J].中国现代应用药学,2010,27(4):304-306. 被引量:4
  • 9朱博,王新鸿,唐伶俐,李传荣.光学遥感图像信噪比评估方法研究进展[J].遥感技术与应用,2010,25(2):303-309. 被引量:31
  • 10胡峰,王中峰,王伟,张洪,高普均.吲哚菁绿清除试验和终末期肝病模型评分[J].肝脏,2012,17(8):551-553. 被引量:8

共引文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部