POCT便携式血糖仪与全自动生化分析仪对比分析

Comparative Analysis of POCT Portable Blood Glucose Meter and Automatic Biochemical Analyzer
下载PDF
导出
摘要 目的 对比分析POCT便携式血糖仪与全自动生化分析仪。方法 配置分布在1~30mmol/L不同血糖浓度的抗凝静脉全血标本5份,用POCT便携式血糖仪检测,每份标本重复检测3次,分离血浆在贝克曼AU5800全自动生化分析仪上进行检测,并对结果进行统计学分析。结果 18台POCT便携式血糖仪与标准靶值的差值及偏倚均在要求范围内,POCT便携式血糖仪的正确度验证合格。相关分析结果显示,AU5800全自动生化分析仪与POCT便携式血糖仪结果显著相关,相关系数r值均大于0.975,结果呈正相关。结论 POCT便携式血糖仪与贝克曼AU5800全自动生化分析仪血糖比对结果一致,POCT便携式血糖仪满足日常血糖监测要求,定期与全自动生化仪进行对比,确保其结果准确性、可靠性。 Objective To regularly monitor consistency of blood glucose results between 18 POCT portable blood glucose meters and Beckman AU5800 automatic biochemical analyzer in our clinical department,to ensure accuracy of POCT portable blood glucose meter detection.Methods The paper configured 5 anticoagulant venous whole blood samples with different blood glucose concentrations ranging from 1 to 30 mmol/L,tested with POCT portable blood glucose meter.Each sample was tested three times,plasma was separated and tested by Beckman AU5800 automatic biochemical analyzer.result were statistically analyzed.Results Difference and bias between 18 POCT portable blood glucose meters and standard target values were within required range,accuracy verification of POCT portable blood glucose meters was qualified.Correlation analysis results showed,there was significant correlation between AU5800 automatic biochemical analyzer and POCT portable blood glucose meter,with correlation coefficients r values greater than 0.975,indicating positive correlation.Conclusion Blood glucose comparison results between POCT portable blood glucose meter and Beckman AU5800 automatic biochemical analyzer are consistent.The POCT portable blood glucose meter can meet daily blood glucose monitoring requirements.Regular comparion with automatic biochemical analyzer can ensure its accuracy and reliability.
作者 康蒙 KANG Meng(Department of Pathology and Laboratory Medicine,Hebei Yanda LU Dao-Pei Hospital,Langfang,Hebei Province,065201)
出处 《智慧健康》 2024年第21期101-104,共4页 Smart Healthcare
关键词 POCT血糖仪 自动生化分析仪 质量管理 对比实验 POCT blood glucose meter Automatic biochemical analyzer Quality management Comparative experiment
  • 相关文献

参考文献7

二级参考文献43

  • 1章秀兰.简易血糖仪使用前培训的重要性[J].检验医学,2004,19(4):327-329. 被引量:46
  • 2樊晓萍,张环生,赵秀风,王国萍.快速血糖仪与全自动生化分析仪检测血葡萄糖结果对比研究[J].国际检验医学杂志,2006,27(3):287-287. 被引量:54
  • 3Clinic and Laboratory Standards Institute.C30-A.Ancillary(bedside)blood glucose testing in acute and chronic care facilites.Approved guideline.Wayne,PA:CLSI,2002.
  • 4ISO. ISO 22870. Point - of - care testing (POCT) - Requirements for quality and competence[ S]. ISO, 2006.
  • 5CLSI. POCT07 - P. Quality Management : Approaches to Reducing Errors at the Point of Care: Proposed Guideline [ S]. CLSI, 2009.
  • 6Guido F,Christina S,Annette B,et al. System accuracy e valuation of 43 blood glucose monitoring systems for self- mornitoring of blood glucose according to DIN EN ISO 15197[J]. J Diabetes Sci Technol,2012,6(5) :1060-1075.
  • 7Chen E,Duh S, Hortin G. Performance evaluation of blood glucose monitoring devieces[J]. Diabetes Technol Ther, 2003,5 (15) : 749-768.
  • 8Halldorsdottir S, Warehal-Windham ME, Wallace JF, et al. Accuracy evaluation of five blood glucose monitoring systems:the North American comparator trial[J]. J Dia- betes Sci Technol,2013,7(5) :1294 1304.
  • 9Brad SK, Laurie G Renee S et al. Evaluation of the im pact of hematoerit and other interference on the accuracy of hospital-based glucose meters[J]. Diabetes Technol T her, 2008,10(2) : 111-120.
  • 10Tang Z, Lee J, Louie R, et al. Effects of different hemato- crit levels on glucose measurements with handheld meters for point-of-care testing[J]. Arch Pathol Lab Med, 2000, 124(16) :1135-1140.

共引文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部