摘要 目的 研究染色法在测试超声波清洗机空化效应中的应用效果。方法 试验分为染色法组和铝箔腐蚀法组,分别使用亚甲基蓝+铜版纸、铝箔进行超声波清洗机空化效应的测试,2组各重复10次,观察2组操作时间、标记点总数、标记点平均面积、标记点分布的标准差。结果 染色法组操作时间(38±1.34)s、标记点平均面积(7.48±1.42)像素、标记点分布的标准差(2.52±0.40)像素,分别小于铝箔腐蚀法组的(215±6.11)s、(115.57±13.98)像素、(3.48±0.71)像素;染色法组标记点总数(1 362±216.19)个,大于铝箔腐蚀法组的(91.10±18.76)个,差异均具有统计学意义(P<0.05)。结论 染色法测试超声波清洗机空化效应的操作时间短,测试记录更为全面。 Objective To study the application effect of dyeing method on measuring cavitation effect of ultrasonic cleaning machine.Methods The experiment was divided into dyeing group and aluminum foil corrosion group.Methylene blue + coated paper and aluminum foil were used to test the cavitation effect of ultrasonic cleaning machine,and the two groups were repeated 10 times each.The operation time,total number of markers,average area of markers and standard deviation of marker distribution of the two groups were observed.Results The operation time(38±1.34) s,the average area of markers(7.48±1.42) pixels and the standard deviation(2.52±0.40) pixels of the distribution of markers in the dyeing group were lower than those in the aluminum foil corrosion group(215±6.11) s,(115.57±13.98) pixels and(3.48±0.71)pixels,respectively.The total number of marks in the dyeing group(1362±216.19) was higher than that in the aluminum foil corrosion group(91.10±18.76),and the differences were statistically significant(P<0.05).Conclusion The dyeing method has a short operation time and more comprehensive test record.
作者 郑伟 桂平 东作哈木 江庆华 ZHENG Wei;GUI Ping;DONG Zuo-ha-mu;JIANG Qing-hua(Affiliated Cancer Hospital of University of Electronic Science and Technology/Sichuan Cancer Hospital,Chengdu 610041,China)
出处 《中国消毒学杂志》 CAS 2024年第11期809-811,共3页 Chinese Journal of Disinfection
关键词 超声清洗机 空化效应 染色 铝箔 sultrasonic cleaner,cavitation effect dyeing aluminum foil
  • 相关文献

参考文献12

二级参考文献93

共引文献97

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部