目的探讨SMART POCT和日立7180检测糖化血红蛋白(HbA1c)的结果相关性。方法首先分别对SMART POCT和日立7180生化仪检测HbA1c进行精密度评价,再随机选取20份样本在两台仪器上分别进行测定,对测定结果进行相关性分析。结果两台仪器的精密度高,所测结果的相关性好,相关系数(r2)为0.981 1。结论 SMART POCT和日立7180检测HbA1c具有很高的一致性。
Objective To analyze the consistency of the detection results of glycosylated hemoglobin (HbA1c) by SMART POCT and Hitachi 7180 biochemical analyzers .Methods Firstly the accuracy of HbA1c detected by the SMART POCT and Hita‐chi 7180 biochemical analyzers were evaluated .And then 20 randomly selected samples were tested by both equipments .A relevance analysis between detected data from two equipments was conducted .Results There was a good correlation between the results of the HbA1c detected by two detecting systems ,with the correlation coefficient of r2 =0 .981 1 .Conclusion There is a good consis‐tency in detection results of glycosylated hemoglobin (HbA1c) detected by SM ART POCT and Hitachi 7180 biochemical analyzer .