期刊文献+

Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography 被引量:3

Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography
下载PDF
导出
摘要 AIM: To compare the anterior segment measurements obtained by rotating Scheimpflug camera(Pentacam) and Scanning-slit topography(Orbscan IIz) in keratoconic eyes.METHODS: A total of 121 patients, 71 males(58.7%)and 50 females(41.3%)(214 eyes) with the diagnosis of keratoconus(KC) were enrolled in this study. Following diagnosis of KC by slit-lamp biomicroscopic examination, central corneal thickness(CCT), thinnest corneal thickness(TCT), anterior chamber depth(ACD),and pupil diameter(PD) were measured by a single examiner using successive instrumentation by Pentacam and Orbscan.RESULTS: There was no significant difference between the two instruments for the measurement of CCT and TCT. In contrast, scanning-slit topography measured ACD(3.46±0.40 mm vs 3.38±0.33 mm, P =0.019) and PD(4.97 ±1.26 mm vs 4.08 ±1.19 mm, P 【0.001) significantly larger than rotating Scheimpflug camera.The two devices made similar measurements for CCT(95% CI:-2.94 to5.06, P =0.602). However, the mean difference for TCT was-6.28(95% CI:-10.51 to-2.06, P =0.004) showing a thinner measurement by Orbscan than by Pentacam. In terms of the ACD, the mean difference was 0.08 mm(95%CI: 0.04 to 0.12, P 【0.001) with Orbscan giving a slightly larger value than Pentacam. Similarly, Orbscan measurement for PD was longer than Pentacam(95% CI:0.68 to 1.08, P 【0.001).CONCLUSION: A good agreement was found between Pentacam and Orbscan concerning CCT measurement while comparing scanning-slit topography and rotatingScheimpflug camera there was an underestimation for TCT and overestimation for ACD and PD. AIM: To compare the anterior segment measurements obtained by rotating Scheimpflug camera(Pentacam) and Scanning-slit topography(Orbscan IIz) in keratoconic eyes.METHODS: A total of 121 patients, 71 males(58.7%)and 50 females(41.3%)(214 eyes) with the diagnosis of keratoconus(KC) were enrolled in this study. Following diagnosis of KC by slit-lamp biomicroscopic examination, central corneal thickness(CCT), thinnest corneal thickness(TCT), anterior chamber depth(ACD),and pupil diameter(PD) were measured by a single examiner using successive instrumentation by Pentacam and Orbscan.RESULTS: There was no significant difference between the two instruments for the measurement of CCT and TCT. In contrast, scanning-slit topography measured ACD(3.46±0.40 mm vs 3.38±0.33 mm, P =0.019) and PD(4.97 ±1.26 mm vs 4.08 ±1.19 mm, P <0.001) significantly larger than rotating Scheimpflug camera.The two devices made similar measurements for CCT(95% CI:-2.94 to5.06, P =0.602). However, the mean difference for TCT was-6.28(95% CI:-10.51 to-2.06, P =0.004) showing a thinner measurement by Orbscan than by Pentacam. In terms of the ACD, the mean difference was 0.08 mm(95%CI: 0.04 to 0.12, P <0.001) with Orbscan giving a slightly larger value than Pentacam. Similarly, Orbscan measurement for PD was longer than Pentacam(95% CI:0.68 to 1.08, P <0.001).CONCLUSION: A good agreement was found between Pentacam and Orbscan concerning CCT measurement while comparing scanning-slit topography and rotatingScheimpflug camera there was an underestimation for TCT and overestimation for ACD and PD.
出处 《International Journal of Ophthalmology(English edition)》 SCIE CAS 2015年第2期275-280,共6页 国际眼科杂志(英文版)
关键词 keratoconus Orbscan Pentacam rotating Scheimpflug camera scanning-slit topography corneal pachymetry central corneal thickness thinnest corneal thickness anterior chamber depth pupil diameter keratoconus Orbscan Pentacam rotating Scheimpflug camera scanning-slit topography corneal pachymetry central corneal thickness thinnest corneal thickness anterior chamber depth pupil diameter
  • 相关文献

同被引文献35

引证文献3

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部