摘要 目的:比较神经反应遥测和行为听力测试两者阈值的差异,探讨神经反应遥测所得阈值对小幼儿言语处理器编程调试的指导价值。方法:比较9例接受Nucleus24型人工耳蜗植入患儿共69个电极的神经反应遥测和行为听力测试的阈值,对所得结果进行t检验。结果:神经反应遥测阈值与行为听力测试阈值比较有统计学差异(P<0.01),前者比后者大约高12电流级(currentlevel)。结论:神经反应遥测阈值可大致预估实际的行为听力测试阈值。
出处 《新医学》 北大核心 2003年第1期32-33,共2页 Journal of New Medicine
  • 相关文献

参考文献2

  • 1[1]Shallop JK, Facer GW, Peterson A. Neural response telemetry with the nucleus CI24M cochlear implant. Laryngoscope, 1999, 109 (11):1755-1759.
  • 2[2]Hughes ML, Brown CJ, Abbas PJ, et al. Comparison of EAP thresholds with MAP levels in the nucleus 24 cochlear implant: data from children.Ear Hear, 2000, 21 (2): 164-174.

同被引文献3

  • 1Hughes ML, Vander Werff KR, Brown CJ, et al. A longitudinal study of electrode impedance, the electrically evoked compound action potential, and behavioral measures in Nucleus 24 cochlear implant users[J]. Ear Hear, 2001, 22:471.
  • 2Dillier N, Lai WK. Measurement of the electrically evoked compound action potential via a neural response telemetry system[J]. Ann Otol Rhinol Laryngol, 2002, 111:407.
  • 3郗昕,洪梦迪,韩东一,黄德亮,杨伟炎.小儿耳蜗植入后电诱发复合动作电位的阈值及其临床应用[J].中华耳鼻咽喉科杂志,2003,38(1):43-46. 被引量:12