摘要
本方法通过正交试验选择优化实验参数 ,测定痕量铅镉的线性范围分别为 4.0 - 2 0 0 μg/l和 2 .0 - 1 2 0 μg/l,加标回收率在 95 - 1 0 7%之间。用此电极测定了矿井水、河水、工业污水中的pb2 +、cd2
A gilded glassy carbon electrode for the voltammetric determination of trace Pb 2+ , Cd 2+ was described. Optimization of experiment parameters is selected by test. The content of lead and cadmium has linear relationship with the stripping current in the range of 4.0-200μg/l and 2.0-100μg/l respectively. The recovery is from 95-107%. Using this electrode, trace levels of Pb 2+ ,Cd 2+ in water was determined.
出处
《河北建筑科技学院学报》
2002年第4期5-6,10,共3页
Journal of Hebei Institute of Architectural Science & Technology
基金
河北省科委资助项目 (0 0 2 76711)