摘要 目的 探讨在线尿素清除率监测(OCM)的临床应用价值及在透析充分性评价中的临床意义。方法 采用4008S血液透析机OCM检测了38例维持性透析患者一次血液透析时的尿素清除指数(Kt/V值),并用尿素动力一室模型计算Kt/V值;根据有、无残余肾功能,将38例分为两组对OCM和计算Kt/V值进行比较。结果 38例患者OCM检测Kt/V值为1.13±0.58,计算Kt/V值为1.18±0.17,具有良好的相关性(r=0.887,P<0.01)。尿素动力模型计算二组的Kt/V值有显著差异(P<0.05),而OCM检测其Kt/V值则无显著差异(P>0.05)。结论 血液透析机OCM可作为检测Kt/V,评价透析充分性的简便、可靠的指标,并可避免透析后尿素反弹和残余肾功能的影响。 Objective To evaluate the clinical efficiency of online clearance measurement (OCM) based on conductivity variation. Methods Kt/V were assessed both by OCM and urea kinetic model in 38 maintenance hemodialysis patients, of which 2 groups were separated according to their residual renal function. Results Kt/V as estimated by OCM were 1.13±0.58, whereas measurements of Kt/V were 1.18±0.17 by urea kinetic model. They showed highly significant linear correlation( r = 0.887, P < 0.01). Measurements of Kt/V by urea kinetic model had significant differences between 2 groups( P < 0.05), while Kt/V as estimated by OCM had no significant differences( P > 0.05). Conclusion The OCM option of the hemodialysis machine provides an accurate tool for online monitoring of urea clearance to evaluate dialysis adequacy.
出处 《中国血液净化》 2003年第1期20-22,共3页 Chinese Journal of Blood Purification
关键词 在线尿素清除率 透析充分性 尿素动力模型 血液透析 Online clearance measurement Dialysis adequacy Urea kinetic model
  • 相关文献

参考文献5

  • 1[1]Daugirlas JT. Second generation logarithmic estimates of single - pool variable volume Kt/V: an analysis of error. J Am Soc Nephrol, 1993,4(5): 1205 - 1213
  • 2[2]Leon JB, Sehgal AR .ldentifying patients at risk for hemodialysis underprescription. Am J Nephrol, 2001,21 (3): 200 - 207
  • 3[3]Hauk M, Kuhlmann MK,Riegel W,et al. In vivo effects of dialysate flow rate on Kt/V in maintenance hemodialysis patients. Am J Kidney Dis,2000,35( 1): 105-111
  • 4李莉,李学旺,刘红,李艳,杨桂琴,赵建成.联机尿素监测仪在临床应用中的准确性及有效性评价[J].中华肾脏病杂志,2000,16(5):320-323. 被引量:7
  • 5[5]Kuhlmann U,Goldau R,Samadi N,et al. Accuracy and safety of online clearance monitoring based on conductivity variation. Nephrol Dial Transplant, 2001 ,16(5): 1053-1058

二级参考文献7

  • 1Schoenfeld P Y,Henry R R,Laird N M,et al.Assessment of nutritional status of the National Cooperative Dialysis Study population[].Kidney International.1983
  • 2Daugirdas JT.Second generation logarithmic estimates of single-pool variable volume Kt/V: an analysis of error[].Journal of the American Society of Nephrology.1993
  • 3Watson PE,Watson ID,Batt RD.Total body water volumes for adult males and females estimated from simple anthropometric measurements[].The American Journal of Clinical Nutrition.1980
  • 4Bosticardo GM,Avalle U,Giacchino F,et al.Accuracy of anon-line urea monitor compared with urea kinetic model and directdialysis quantification[].ASAIO Journal.1994
  • 5Depner TA,Deshaviah PR,Ehben JP,et al.Multicenter clinicalvalidation of an on-line monitor of dialysis adequacy[].Journal of the American Society of Nephrology.1996
  • 6Canaud B,Bosc JY,Leblanc M,et al.Evaluation of high-fluxhemodilafiltration efficiency using an on-line urea monitor[].American Journal of Kidney Diseases.1998
  • 7Garred LJ,Barichello DL,Canaud BC,et al.Simple equations forprotein catabolic rate determination from pre dialysis and postdialysis blood urea nitrogen[].ASAIO Journal.1995

共引文献6

同被引文献76

引证文献13

二级引证文献100

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部