期刊文献+

有限角CT重构压缩算法及仿真分析

The Analysis of Squashing Algorithm for Limited Angle Projection Reconstruction by Simulation
下载PDF
导出
摘要 研究了有限角 CT重构压缩算法中有限角度和投影幅数对重构结果的影响 .通过仿真程序 ,对不同的有限角度和投影幅数下的压缩重构结果进行分析和比较 ,得出不同的有限角度和投影幅数下重构质量有很大的不同 .当缺失角大于等于 3 0°时 ,重构质量很差 . The effect of the limited angle and the number of projections in squashing algorithm for limited angle projection reconstruction are studied. Via simulations, the reconstructed images under different limited angle and the different number of projections by the squashing algorithm are analysed and compared. It is shown that the quality of the reconstructed image has large difference between the different limited angle and the different numbers of projection. When the lost angle is equal to 30° or larger than, the quality of the reconstructed image are deteroited badly. It illuminates that the squashing algorithm is valid only in a narrow fields.
出处 《测试技术学报》 2002年第4期235-237,共3页 Journal of Test and Measurement Technology
关键词 计算机断层成像 有限角 重构 压缩算法 投影幅数 computer tomography limited view angle reconstruction algorithm
  • 相关文献

参考文献3

  • 1王召巴,金永.高能X射线工业CT技术的研究进展[J].测试技术学报,2002,16(2):79-82. 被引量:16
  • 2Reeds J A; Shepp L A. Limited Angle Reconstruction in Tomography via Squashing[J]. IEEE Trans Med Imag. 1987; MI-6(2): 89-97.
  • 3庄天戈. CT 理论与算法[M]. 上海: 上海交通大学出版社; 1992; 30-60.