石墨炉原子吸收光谱法测定磷时基体改进剂的应用研究 被引量:6

Application of Matrix Modifier to the Determination of Phosphorus by Graphite Furnace Atomic Absorption Spectrometry
下载PDF
导出
摘要 本文阐述几种基体改进剂测定磷的效果,发现Pd/Ca混合基体改进剂对测定植物和动物组织中磷最理想,利用氘灯扣背景,其特征量为8.2ng,吸光度线性范围为0~0.220,磷含量为0.144%和1.35%时,相对标准偏差为2%左右,精密度好,抗干扰能力强,测定茶树叶和贻贝中磷比较理想。 The effect was presented for the determination of phosphorus with matrix modifier by graphite furnace atomic absorption spectrometry. It was found from results that Pd/Ca as a matrix modifier was more reasonable to determine phosphorus in animal and plant samples. The linear relationship between the absorbance and the concentration of phosphorus in the range of 0~20 μg/ml was obtained. The results were in good agreement with the reference and standard values.
作者 姚金玉 杨琼
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1992年第2期183-185,共3页 Chinese Journal of Analytical Chemistry
关键词 石墨炉 基体改进剂 FAA 测定 Phosphorus, Matrix modifer, Graphite furnace atomic absorption spectrometry