期刊文献+

超声雾化进样微波诱导等离子体原子发射光谱法测定痕量磷的研究 被引量:4

Determination of Phosphorus by Microwave Induced Plasma Atomic Emission Spectrometry Using an Ultrasonic Nebulizer
下载PDF
导出
摘要 本文提出了一种测定水溶液样品中痕量磷的微波诱导等离子体原子发射光谱法(MIPAES)。样品经超声雾化并去溶后引入等离子体。考查了Ar和He MIP中磷的发射光谱特性和一些实验参数对磷发射强度的影响。用Ar和He作工作气体时,检出限分别为0.03和0.0045μg/ml,方法的相对标准偏差小于5%。 The determination method of phosphorus by microwave induced plasma (MIP) atomic emission spectrometry was investigated. Aqueous samples were introduced with an ultrasonic nebulizer and a desolvation system. The emission spectroscopic behavior of phosphorus in Ar and He MIP was investigated. The influence of some experimental conditions on emission intensity of phosphorus was examined. The detection llmits for phosphorus in Ar and He as the carrier gas were 0.03 and 0.0045μg/ml, respectively. The relative standard deviation was<5%.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1992年第9期1065-1068,共4页 Chinese Journal of Analytical Chemistry
基金 国家自然科学基金
关键词 超声雾化 MIP-AES Microwave induced plasma atomic emission spectrometry Ultrasonic nebulization Phosphorus.
  • 相关文献

参考文献4

二级参考文献2

  • 1金钦汉,Proc 2nd Beijing Conf and Exhib on Instrum Analysis,1987年
  • 2金钦汉,光谱学与光谱分析,1985年,5卷,1期,36页

共引文献15

同被引文献40

引证文献4

二级引证文献32

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部