摘要
本文建立了痕量芴、苊、蒽和(艹北)的同时偏振-同步荧光分析法.该法能有效地抑制同步扫描过程中散射光的影响,从而提高测定的灵敏度.芴、苊、蒽和(艹北)的检测限分别为(ng/mL):1.1、10.0、1.8、0.26,相对标准偏差不大于4.2%.
The present paper contains a new method of polarization-synchronization spectrofluo-rimetry for the simultaneous determination of trace fluorene.acenaphthylene,anthracene and perylene.The use of this technique can effectively suppress the effect of scatteredlight during synchronous scanning,and hence enhance sensitivity.The detection limits of fluorene,acenaph-thene anthracene and perylene were(ng/mL):1.1,10.0,1.8 and 0.26,respectively.The relative standard deviation was less than 4.2%.
出处
《高等学校化学学报》
SCIE
EI
CAS
CSCD
北大核心
1992年第10期1214-1216,共3页
Chemical Journal of Chinese Universities
关键词
芴
苊
蒽
Bei
荧光分析
偏振分析