复合材料人工髋关节的实验研究 被引量:6

Experimental Study on Polysulfone Composite Hip Prosthesis
下载PDF
导出
摘要 目的:应力遮挡引起的股骨上段骨吸收和远期松动是影响人工髋关节远期疗效的主要原因之一。本实验对弹性模量接近皮质骨的聚砜复合材料及其人工髋关节进行了研究。方法:用14只成年狗行右侧髋关节置换术,植入聚砚-碳纤维-羟基磷灰石复合材料假体。分术后当日、1月、3月、6月和9月组处死动物。通过单光子骨密度仪,X线摄片计算机辅助分析,光镜和扫描电镜分析,对假体植入后股骨上段骨重建变化和骨假作界面结合进行研究。结果:显示复合材料假体应力遮挡轻,假体植入侧骨密度变化与对照组相比,无骨吸收。骨假体界面结合满意,1个月结合面积(45.61±2.75)%、3个月(50.26±2.19)%、6个月(51.24±1.07)%、9个月(54.02±4.08)%。结论:聚砜复合材料假体有良好的力学相容性和生物相容性,是一种有前途的新型假体。 Objective: Cortical bone atrophy of proximal femur and prosthesis loosening secondary to stress shielding are major complications in long term result of total hip arthroplasty (THA). To solve the problem, authors developed a polysulfone composite hip prosthesis. Methods: The right hip of fourteen adult dogs were replaced with polysulfone-carbonfiber hydroxylapatite (HA) composite hip prosthesis. Animals were sacrificed on the operation day, one month, three months, six months and nine months after operation and bone remodeling of proximal femur and bone prosthesis interface integration were analyzed by single photon absorptiopmetry(SPA), computer assisted X-ray films analysis, light microscopy, fluorescent microscopy and scanning electron microscopy for histological and ultrahigh structure. Results:The composite hip prosthesis has little stress shielding and satisfaction in bone prosthesis interface integration. Bone-prosthesis integrative contact was (45. 61 ± 2. 75) % in one month, (50.26±2.19)% in three months,(51.24±1.07)% in six months and (54.02±4.08)% in nine months.Conclusion:The composite hip prosthesis has satisfactory mechanical compatibility and biocompatibility. It is a promising hip prosthesis.
出处 《中国矫形外科杂志》 CAS CSCD 2003年第3期214-216,共3页 Orthopedic Journal of China
关键词 实验研究 人工髋关节 复合材料 聚砜 羟基磷灰石 碳纤维 Hip prosthesis Composite material Polysulfone Hydroxylapatite Carbofiber
  • 相关文献

参考文献5

  • 1[1]Sumner DR,Gallant JO.Determinants of stress shielding:Design versus materials versus interface[J].Clin Orthop, 1991,274:202~212.
  • 2[2]Engh CA,Massin P,Suthers KE.Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components[J].Clin Orthop,1990,257:107~128.
  • 3[3]Huiskes R,Weinans H,Rietbergen B.The relationship between stress shielding and bone resorption around total hip stem and the effects of flexible materials[J].Clin Orthop,1992,274:124~134.
  • 4[4]Sumner DR,Turner TM,Urban RM,et al.Experimental of bone remodeling in total hip arthroplasty[J].Clin Orthop,1992,276:83~90.
  • 5[5]Rotem A.Effect of implant material properties on the performance of a hip joint replacement[J].J Med Eng Technol (ENGLAND),1994,18(6):208~214.

同被引文献93

引证文献6

二级引证文献41

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部