二级标准血液分析仪的溯源与质量控制

下载PDF
导出
摘要 目的保证全血细胞计数(complete blood count,CBC)检测结果的溯源性和准确性。方法参加卫生部临检中心建立的血细胞分析溯源体系,向卫生部临检中心参考实验室进行血细胞分析的溯源和比对。采用二级标准血液分析仪,用配套或卫生部临检中心参考实验室提供的定值新鲜血校准,并使用与参考实验室相同批号的全血质控物、定值乳胶微粒质控物(Cell- check-400)和氰化高铁血红蛋白标准液(HISTAN)作为比对物质进行质量控制,每月将全血质控物检测结果与参考实验室同期测定而提供的靶值进行比对,Cellcheck-400和HISTAN检测结果与其定值比对。结果全血质控物检测结果与靶值的相对偏差:WBC±4.5%内,RBC±5.5%内,HGB±2.5%内,HCT±4.0%内,PLT±6.8%内;Cellcheck-400和HISTAN检测结果与其定值相对偏差:RBC±3.5%内,MCV±4.3%内,HCT±4.0%内,HGB±1.8%内。结论二级标准血液分析仪检测结果的可比性良好,其溯源模式和质量控制方案保证了二级标准血液分析仪检测结果的溯源性和准确性。
出处 《浙江检验医学》 2005年第4期21-22,28,共3页 Zhejiang Journal of Laboratory Medicine
  • 相关文献

参考文献9

二级参考文献31

  • 1国家质量技术监督局计量司.测量不确定度评定与表示指南[M].北京:中国计量出版社,2000.62-124.
  • 2[1]International Committee for Standardization in Hematology;Expert p anel on cytometry.Reference method for the enumeration of erythrocytes and leuk ocytes[J]. Clin Lab Heamat,1994,16∶131-138.
  • 3[2]International Committee for Standardization in Hematology;Expert panel o n hemoglobinometry.Recommendations for reference method for hemoglobinometry in human blood(ICSH standard 1986)and specifications for international hemoglobincy anide reference preparation 3rd[J].Clin Lab Haematol,1987,9∶73-79.
  • 4[3]NCCLS.Precedure for Determining Packed Cell Volume by the Microhematioc r it Method-Second Edition;Approved Standard[S].NCCLS Document H7-A2, Vol.1 3 No 9,1993,13(9).
  • 5European Union. Directive 98/79/EC of the european parliament and of the council of 27 october 1998 on in vitro diagnostic medical devices.Official Journal of the European Communities, 1998, L331 : 1-37.
  • 6International Organization for Standardization. In vitro diagnostic medical devices-measurement of quantities in samples of biological originmetrological traceability of values assigned to calibrators and control materials. ISO/DIS 17511, International Organization for Standardization,Geneva,2000.
  • 7International Organization for standardization. In vitro diagnostic medical devices-measurement of quantities in samples of biological originmetrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials. ISO/DIS 18153,International Organization for Standardization, Geneva, 2000.
  • 8International Organization for Standardization. General requirements for the competence of testing and calibration laboratories. ISO/IEC 17025,International organization for standardization, Geneva, 1999.
  • 9International Organization for Standardization. Quality management i0n the medical laboratory. ISO/FDIS 15189, International Organization for Standardization, Geneva, 2000.
  • 10Siekmann L. Traceability of calibration for IVD industry: Process for eredentialing reference materials, reference methods and reference laboratories in europe. In: proceedings of the workshop on measurement traceability for clinical laboratory testing and in vitro diagnostic test systems. National Institute of Standards and Technology, Gaithersburg MD, 2000. 85-101.

共引文献192

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部