摘要 本文研究了ICP-AES氢化物发生法测定生物样品中痕量硒的方法。对氢化物发生条件和样品消化过程进行了选择,并使之达最佳化。方法检出限达0.09ng/ml。 A method is studied for the determination of selenium in biological samples by continuous hydride generation ICP-AES. A variety of sample dissolution procedures and hydride generation reaction conditions were evaluated in attempt to establish optimal conditions. Detection limit is 0.09ng/ml for Se. Results of analyses of standard reference materials and many biological samples demonstrate the applicability of the technique.
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 1992年第5期67-70,54,共5页 Spectroscopy and Spectral Analysis
基金 国家自然科学基金
关键词 生物样品 ICP-AES 氢化物发生 Selenium Biological sample ICP-AES Hydride generation