联机清除率监测器在复用透析器效果监测中的作用评价 被引量:2

Assessment of the role of online clearance-rate monitor on monitoring the effect of reused dialyser
原文传递
导出
摘要 目的 对联机清除率监测器 (OCM)在复用透析器效果监测中的作用进行评价。方法  2 0例接受血液透析患者 ,分别在透析器首次使用和复用第 5次时用OCM和常规取血法测定尿毒清除指数 (Kt/V) ,同时测定总透析器容量 (TCV)。结果 透析器首次使用与复用第 5次时 ,OCM和常规取血法测定Kt/V值均差异无显著性 (P >0 0 5 ) ,而且两种方法的Kt/V测定结果直线相关分析均有显著意义 (P <0 0 1) ;TCV测定结果与OCM法和常规取血法Kt/V测定结果直线相关分析均无显著意义 (P >0 0 5 )。结论 OCM法监测复用透析器的Kt/V使用方便、效果可靠 ,可代替常规取血法Kt/V测定 ;TCV测定不能准确反映Kt/V 。 Objective To assess the role of online clearance rate monitor(OCM) on monitoring the effect of reused dialyser.M20 patients on hemodialysis were recruited in this study.Urea clearance index (Kt/V) was measured by the methods of OCM and traditional blood analysis on the first time used dialyser and the fifth time used dialyser respectively,while total column volume(TCV) was also determined.Results For either the first time used dialyser or the firth time used one,there was no sigificant difference between the OCM results and blood analysis results(P>0 05),while there was significant linear correlation between the two Kt/V detection methods(P<0 01).For the fifth time used dialyser,there was no significant correlation between TCV and OCMs Kt/V (P>0 05),nor between TCV and blood analysiss Kt/V. Conclusion OCM can be used to monitor Kt/V of reused dialyser conveniently and reliably,with the possibility to replace blood analysis.TCV cant reflect Kt/V accureately,so it cant act as the sole index of the monitoring effect of reused dialyser. [
出处 《中华急诊医学杂志》 CAS CSCD 2003年第2期84-85,共2页 Chinese Journal of Emergency Medicine
关键词 肾脏疾病 血液透析疗法 联机清除率监测器 尿毒清除指数 复用透析器 Online clearance rate Usea clearance index Reused dialyser
  • 相关文献

参考文献1

共引文献1

同被引文献14

  • 1谢华,林洪丽,李平,陈丽光,王松岩,冯青.联机尿素清除率监测和常规采血计算Kt/V对血液透析充分性评估的对照研究[J].中国血液净化,2004,3(11):587-589. 被引量:22
  • 2Wuepper A,Tattersall J, Kraemer M, et al. Detemination of urea distribution volume for Kt/V assessed by conductivity monitoring [ J ]. Kidney Int,2003 ; 64 (6) : 2262 - 2271
  • 3Lindsay RM, Bene B, Goux N. Relationship between effective ionic dialysance and in vivo urea clearance during hemodialysis[ J ] .Am J Kidney Dis,2001 ;38(3) :565 -574
  • 4梅长林,叶朝阳,赵学智.实用透析手册[M].北京:人民卫生出版社,2004:166~179.
  • 5WOLFE R A,ASHBY V B,DAUGIRDAS J T,et al.Body sized,dose of hemodialysis.and mortality[J].Am J Kidney Dis,2000,35:80-88.
  • 6MURTHY B V,SUNDARAM S,JABER B L,et al.Effect of formaldehyde/bleach reprocessing on in vivo performances of high -efficiency cellulose and high-flux polysulfone dialyzers[J].J Am Soc Nephrol,1998,9:464-472.
  • 7SCHNEDITZ D,MARTIN K,KRAMER M,et al.Effect of controlled extracorporeal blood cooling on ultrafiltmtion-induced blood volume changes during hemodialysis[J].J Am See Nephml,1997,8:956-964.
  • 8约翰.T,多格达斯,彼得.G.布莱克,等.透析手册[M].2版.沈阳:辽宁教育出版社,2004:10-19.
  • 9TIAO L H,CHIEN T L.Low serum albumin and high ferritin levels in chronic hemodialysis patients with major depression[J].Psychiatry Research,2007,8(15):277-280.
  • 10QURESHI A R,ALVESTRAND A,DIVINO FILHO J C,et al.Inflammation,malnutrition,and cardiac diseases predictors of mortality in hemodialysis patiens[J].J Am Soc Nephrol,2002,13(S1):26-28.

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部