摘要 进行了用日立180-80石墨炉原子吸收光谱仪测定食品中镉的分析条件的优化研究。结果表明,食品中镉测定的最佳分析条件为:灯电流7.5~12.5mA(样品的镉含量较低时选择较小电流,反之则选较大电流);干燥、灰化、原子化、净化的温度和时间分别为120℃、15~20s,500~550℃、20s,1500℃、5s,2600℃、3s。 Studies were made on optimization of conditions for determining the c ad mi um in food by Hitachi 180-80GFAAS.The results showed that the best con ditions for determining the cadmium in food were electric currents through the l ight 7.5-12.5mA(For sam ples with low cadmium contents,low electric cu rrents were chosen;and for samples with high cad mi um contents,high electric currents were chosen).The temperatures and times for dry ing,ash ing,atomizi ng,purifying were120℃,15-20s;500-550℃,20s;1500℃,5 s;2600℃,3s;respectively.
作者 周聪
出处 《热带农业科学》 2002年第6期24-28,共5页 Chinese Journal of Tropical Agriculture
关键词 食品检验 石墨炉原子吸收光谱仪 测定 分析条件 优化 instrument analysis method GFAAS cadmium food chemistry
  • 相关文献

参考文献4

  • 1国家质量监督检验检疫总局.中华人民共和国国家标准--农产品安全质量无公害蔬菜安全要求.GB 18406.1-2001.北京:中国标准出版社,2001.4
  • 2国家质量监督检验检疫总局.中华人民共和国国家标准--农产品安全质量无公害水果安全要求.GB18406.2-2001.北京:中国标准出版社,2001.12
  • 3杨惠芬 李明元.食品卫生理化检验标准手册[M].北京:中国标准出版社,1997.115.
  • 4中华人民共和国卫生部.中华人民共和国国家标准--食品中铅的测定方法.GB/T5009.15-1996.北京:中国标准出版社,1996.1-3