摘要 实验测得φ40mm×40mmBGO在GDB50L和GDB44D光电倍增管上的时间分辨率为1.9,3.4ns(对^(60)Co),φ76mm×100mmBGO在GDB76F上的时间分辨率为6.3ns。用蒙特卡罗方法研究了光电倍增管对探测器时间性能的影响,预计φ76mm×100mmBGO在3英寸快管上的时间分辨率可在2.5ns以下。 The time resolutions (for 60Co source) of the detectors composed of $40 mmX40 mm BGO+GDB50L,$40mmX40 mm BGO+GDB44D and $76 mmXIOO mm BGO+GDB76F were measured to be 1.9 ns, 3.4 ns and 6.3 ns> respectively. The influence of the photomultiplier tubes on the timing properties of the detectors was studied with the Monte Carlo Simulation.The calculation showed that the time resolution of the $76 mm, 100 mm B<fO a fast 3' -diameter photomutiplier tude could be blow 2.5 ns.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 1992年第4期220-224,共5页 Nuclear Electronics & Detection Technology
关键词 BGO 时间分辨 探测器 圆柱体型 BGO, Time resolution, Monte Carlo method
  • 相关文献

参考文献2

  • 1叶沿林,刘昕,江栋兴,卢希庭,刘洪涛,林俊松,周长春.φ76mm×100mm BGO探测器性能研究[J].高能物理与核物理,1992,16(4):297-300. 被引量:3
  • 2张天保,唐孝威,张勤建.国产锗酸铋(BGO)闪烁体定时性能的研究[J]高能物理与核物理,1983(06).

二级参考文献4

  • 1朱国义,高能物理与核物理,1990年,14卷,200页
  • 2王--仔,高能物理与核物理,1986年,10卷,1页
  • 3张天保,高能物理与核物理,1983年,7卷,674页
  • 4叶沿林,核电子学与探测技术