摘要
用研制的密封式氙三相点容器№1和№2(不同纯度)复现了氙三相点。其复现性为0.05mk,不确定度为0.6mk。国际比对表明,所测氙三相点分别比用澳大利亚容器测量的氙三相点温度值高0.1mk和1.2mK。
The triple point of natural xenon was realized with a China-made sealed cell at the NIM. The reproducibility of realization is about 0.05 mK. The uncertainty of experimental data is 0.6 mK. The results are in good agreement with NML's cell within 0.1 mK and 1.2 mK respectively.
出处
《计量学报》
CSCD
1989年第4期249-254,共6页
Acta Metrologica Sinica