摘要 国产微机量热仪不同的点火方式对煤的发热量测值有着显著影响 .通过与国标方法对比分析 ,总结得出搅拌 5min再点火是较合理的点火方式 ,从而解决了国产微机量热仪测值偏高的问题 ,为煤炭以质计价提供了公正可靠的计量方法 . Different fire lighting ways of microcalorimeters made in our country have marked effect on the value of heat that coal sends out.Compared with the standard way,it is a better way to light the coal sfter it has been stirred for 5min. It is a more reasonable way. Thus,we settled the problem that the heat value we survey with a calorimeter made in China is often larger.Therefore we can offer you a reasonable and dependable method of determining a coal price according to its quality.
出处 《周口师范学院学报》 CAS 2003年第2期26-27,72,共3页 Journal of Zhoukou Normal University
关键词 煤炭 发热量 微机量热仪 点火控制方式 测量方法 计量方法 calorimeter fire lighting control heat that coal sends out
  • 相关文献

参考文献5

  • 1陈冬英.用国标测试发热量的一点体会[J].燃料纵横,1998,10(2):22-23.
  • 2陈文敏 张自勋.煤化学基础[M].北京:煤炭工业出版社,1995..
  • 3李英华.煤质分析应用技术指南[M].北京:中国标准出版社,2001..
  • 4.GB/T213-1996.煤的发热量测定方法[S].北京:中国标准出版社,1996..
  • 5.微机量热仪技术说明书[Z].长沙友欣设备有限公司,2000..

共引文献9

同被引文献14

  • 1李英华.煤质分析应用技术指南.第二版,北京:中国标准出版社.2009.
  • 2GB/T213-2008煤的发热量测量方法.
  • 3蒋伟.基于S3C2410的恒温式自动量热仪的设计:[学位论文].郑州:郑州大学.2009.
  • 4李平.基于单片机的恒温式自动量热仪的研究与设计:[学位论文].郑州:郑州大学,2011.
  • 5李英华.煤质分析心川技术指南[M].北京:中国标准出版社,2009.
  • 6李平.綦于单片机的恒温式自动最热仪的研究与没计[D].郑州:郑州大学硕士论文,2011.
  • 7李平.基于单片机的恒温式自动量热仪的研究与设计[D].郑州:郑州大学,2011.
  • 8胡彪,孟志强,胡如龙.PCI总线技术在量热仪系统上的应用[J].仪表技术与传感器,2008(6):41-43. 被引量:4
  • 9单勇,徐广立.智能自动量热仪的应用[J].山东煤炭科技,2009,27(1):59-60. 被引量:5
  • 10蒋伟,赵书俊,徐航,段志峰.基于S3C2410的恒温式自动量热仪设计[J].现代电子技术,2010,33(15):203-205. 被引量:4

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部