期刊文献+

TIP-300插入式耳机与TDH-50P耳罩式耳机的耳间衰减比较 被引量:2

Comparison of Inter-aural Attenuations between the TIP-300 insert Earphone and the Telephonics TDH-50P Supra-aural Earphone
下载PDF
导出
摘要 目的 比较TIP - 30 0插入式耳机耳间衰减 (insertearphoneinterauralattenuation ,IEIA)与TDH - 5 0P耳罩式耳机耳间衰减 (supra -auralearphoneinterauralattenuation ,SEIA)的差异 ,为TIP - 30 0插入式耳机的临床应用提供参考依据。方法 利用GSI6 1临床听力计、TIP - 30 0插入式耳机和TDH - 5 0P耳罩式耳机 ,对一组单耳全聋而另一耳听力正常者 35人 (男 13人 ,女 2 2人 )进行纯音气导的耳间衰减测试。结果 TIP - 30 0插入式耳机与TDH - 5 0P耳罩式耳机组间耳间衰减有显著性差异 ;其中TIP - 30 0插入式耳机组内的某些频率之间耳间衰减也有显著性差异。结论 在低中频测听范围 ,TIP - 30 0插入式耳机的耳间衰减比TDH - 5 Objective To afford a clinical reference for applications of TIP-300 insert earphone in accordance with its difference in inter-aural attenuation from the telephonics TDH-50P supra-aural earphone.Methods Detections of inter-aural attenuation for pure tone air conduction were performed on the enrolled individuals combined with unilateral anakusis and one normal ear (male and female were 13 and 22 respectively) by GSI61 audiometer?TIP-300 insert earphone or telephonics TDH-50Psupra-aural earphone.Results Inter-aural attenuation differences were significant ( P <0.001) between the TIP-300 insert earphone group and the TDH-50P supra-aural earphone group when the frequence ranged from 125 to 8 000 Hz.Additionally,attenuations at different frequencies within the TIP-300 group differed significantly.Conclusion TIP-300 insert earphone had a much stronger inter-aural attenuation than telephonics TDH-50 P supra-aural earphone at low or medium frequency spectrum audiometry.
出处 《听力学及言语疾病杂志》 CAS CSCD 2003年第3期180-181,共2页 Journal of Audiology and Speech Pathology
关键词 TIP-300插入式耳机 TDH-50P耳罩式耳机 耳间衰减 比较 耳聋 临床应用 Audiometry Insert earphone Inter-aural attenuation
  • 相关文献

参考文献6

  • 1黄选兆 汪吉宝.实用耳鼻咽喉科学[M].北京:人民卫生出版社,1999.98-99,242-243.
  • 2.GB/T16403-1996声学测听方法:纯音气导和骨导听阈基本测听方法[M].北京:中国标准出版社,1997年6月.7~9.
  • 3Smith DW, Olszyk VB. Auditory behavioral thresholds for Japanese macaques using insert earphones[J]. Division of Otolaryngology- Head and Neck Surgery, 1997,41: 323.
  • 4Hall JW, Grose JH. The effect of conductive hearing loss on the masking-level difference: insert versus standard earphones[J]. J Acoust Soc Am,1994, 95:2652.
  • 5Munro KJ,Agnew N. A comparison of inter- aural attenuation with the Etymotic ER - 3A insert earphone and the Telephonics TDH - 39 supra - aural eaxphone[J]. Br J Audiol, 1999,33: 259.
  • 6Arnold S, Burkard R. Studies of interaural attenuation to investigate the validity of a dichotic difference tone response recorded from the inferior colliculus in the chinchilla[J]. J Acoust Soc Am,2000, 107:1541.

共引文献79

同被引文献11

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部