部分二氧化碳重复吸入法与温度稀释法测定心排血量的相关性研究 被引量:2

The correlation between partial CO_2 rebreathing and thermodilution for cardiac output monitoring
下载PDF
导出
摘要 目的 研究部分二氧化碳重复吸入法和温度稀释法监测心排血量 (CO)的相关性。方法 重症监护病房 (ICU) 1 5例危重患者 ,在呼吸机辅助控制通气条件下 ,采用美国无创心肺功能监测仪 (NICO)用部分二氧化碳重复吸入法 (RB)连续监测无创心排血量 (RBco) ,并同时采用温度稀释法测定心排血量 (TDco) ,两种方法所测CO值进行自体配对相关性研究。结果 配对t检验示两种方法测定的CO值无显著性差异 (P >0 0 5 ) ,相关回归分析示相关系数为 0 90 ,回归方程为RBco =0 81TDco +1 2 5 ,偏离为 0 2 0 ,精确度为 0 78。结论 部分二氧化碳重复吸入是一种简便、快捷、连续、无创的CO测定方法 ,该法与温度稀释法有良好的相关性 ;在无条件行肺动脉漂浮导管插入有创心功能监测的医院 。 Objective To evaluate the correlation between the methods of partial CO 2 rebreathing and thermodilution for cardiac output measurement.Methods Fifteen critical patients receiving mechanical ventilation were studied.Each patient was instrumented with a pulmonary artery thermodilution(TD)cathether and a partial CO 2 rebreathing(RB)ventilator circuit.Cardiac output was measured simultaneouly by RB and TD methods.Paired t test and regression statistics were used to compare the degree of agreement between concurrent measurements of TDco and RBco.Results Paired t test analysis revealed no significant difference between RBco and TDco measurements.Regression analysis gave an r= 0 90 with a slop of 0 81 and an offset of 1 25L/min.Bland Altman analysis resulted in a bias of 0 20 L/min with a precision(ISD)of 0 78L/min.Conclusion There is a high correlation between RBco and TDco measurements.Partial CO 2 rebreathing is a simple noninvasive and automatic method for CO measurement. [
作者 刘琼 黄纪坚
出处 《临床麻醉学杂志》 CAS CSCD 2003年第7期390-391,共2页 Journal of Clinical Anesthesiology
关键词 部分二氧化碳重复吸入法 温度稀释法 测定 心排血量 相关性研究 Cardiac output Partial CO 2 rebreathing Thermodilution
  • 相关文献

参考文献7

  • 1杨跃武.监测呼出和部分重吸入气体中CO_2测心排血量[J].国外医学(麻醉学与复苏分册),2000,21(4):206-207. 被引量:2
  • 2Osterlund B,Gedeon A,Krill P,et al. A new method of using gas exchange measurement for the noninvasive determination of cardiac output, clinical experiences in adults following cardiac surgery. Acta Anaesthesiol Scand,1995,39:727.
  • 3Bailey PL, Haryadi DG, Orr JA, et al. Partial CO2 rebreathing Fick technique for noninvasive measurement of cardiac output. Anesthesia Analgesia, 1998,86:124.
  • 4Watt RC, Loeb RG,Orr J. Comparison of a new non-invasive cardiac output xeehnique with invasive bolus and continuous thermodilution. Anesthesiology, 1998,89 : 536.
  • 5Maxwell RA,Gibson JB,Slade JB,et al. Noninvasive cardiac output by partial CO2 rebreathing after severe chest traum. Traum,2001,51.849.
  • 6Odenstedt H, Stenqvist O, Lundin S. Clinical evaluation of a partial CO2 rebreathing technique for cardiac output monitoring in critically ill patients, Acta Anaesthesiol Scand,2002,46 : 152.
  • 7Johnson KB, Haryadi DG,JA Orr,et al. Influence of pulmonary edema on noninvasive measurement of cardiac output using partial CO2 rehreathing in a canine model.Anesthesiology, 1998,89 : 535.

二级参考文献9

  • 1Watt RC et al. Anesthesiology . 1998
  • 2Jopling MW et al. Anesthesiology . 1998
  • 3Guzzi L. et al. Anesthesiology . 1998
  • 4Haryadi DG et al. Anesthesiology . 1998
  • 5Kück K et al. Anesthesiology . 1999
  • 6Kück K et al.SCCM,1999;abstract No[].D.
  • 7Orr J et al. Critical Care Medicine . 1999
  • 8Johnson KB et al. Anesthesiology . 1998
  • 9Haryadi DG et al. Anesthesiology . 1998

共引文献1

同被引文献13

  • 1Horst H M, Obeid FN, ViJ D, et al. The risks of pulmonary arterial catheriztion[J]. Surg 蒙古 Gynecol Obster, 1984,159:229 - 232.
  • 2Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement[J]. Lancet 1986, 1 : 307 - 310.
  • 3Bland MJ, Altman DG. Calculating correlation coefficients with repeated observation: Part Ⅰ - Correlation within subjects[J]. Br Med J, 1995, 310-446.
  • 4Binder JC, Parkin WG. Non - invasive cardiac output determination: comparison of a partial - rebreathing technique with thennodilution[ J]. Anaesth Intensive Care, 2001,29( 1 ) : 19 - 23.
  • 5Neuhauser C, Muller M, Brau M, et al. Partial CO2 rebreathing technique versus thermodilution: measurement of cardiac output before and after operation extracorporeal circulation [ J ]. Anesthesiology, 2002, 51 (8) : 625 - 633.
  • 6Odenstedt H, Stenqvist O, Lundin S. Clinical evaluation of a partial CO2 rebreathing tectmique for cardiac output monitoring in critically ill patients[J]. Acta Anaesthesiol Scand,2002,46: 152.
  • 7Johnson KB, et al. Influence of pulmonary edema on noninvasive measurement of cardiac output using partial CO2 rebreathing in a canine model. Aneathesiology, 1998, 89(3A): A535.
  • 8Haryadi DG, et al. Evaluation of a partial CO2 rebreathin Fick technique for measurement of cardiac output. Aneathesiology,1998, 89(3A), A534.
  • 9Kuck K, et al. Evaluation of the NICO partial CO2 rebreathing cardiac output monitor in spontaneously breathing animals.Aneathesiology, 1999, 91(3A) : A560.
  • 10杭燕南 庄心良 蒋豪.血管活性药和正性肌力药[A]..当代麻醉学:第1版[C].上海:上海科学技术出版社,2000.367-388.

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部