摘要 调研了各类钍射气(220Rn)测量仪的刻度方法和存在的问题。针对各类220Rn标准源的析出百分率常小于100%而有待准确测定这一关键问题,建立了流气闪烁室绝对测量方法,并用活性炭g谱法作了验证,解决了准确测定220Rn标准源析出率的问题,为提高220Rn测量仪刻度的准确性创造了条件。 The calibration of various kinds of thoron monitor has been investigated. A main problem lies in that the emanation percentage of various kinds of thoron standard source is always less than 100%. In order to determine the emanation percentage accurately, an absolute measurement method with airflow-through scintillation cell was established, and its feasibility was verified by a comparative measurement with the activated charcoal-g spectrometer method. This will make the calibration accuracy of thoron monitors rising.
作者 丘寿康
出处 《核技术》 CAS CSCD 北大核心 2003年第7期561-563,共3页 Nuclear Techniques
关键词 ^220Rn测量 刻度 流气法 Thoron, Measurement, Calibration, Airflow method CLC O613.16
  • 相关文献

参考文献3

  • 1周剑良 邱小平 于涛 等.环境Rn/Tn的测量与评价国际学术研讨会论文汇编[J].湖南衡阳,2002,5:143-48.
  • 2Keane A T, Brewster D IL Health Physics, 1983, 45(3):801-805.
  • 3Kuchta D R, Rundo J, Holtzman R B. Biological and Environmental Effects of low-level Radiation, Vol. II,Proc. Symp, Chicago (Vienna: IAEA), Nov 1975.183--196.