Zebris测量脊柱侧凸的可靠性及重复性研究 被引量:5

Reproducibility and reliability of a new ultrasonography-based system for measuring scoliosis
下载PDF
导出
摘要 目的:研究一种新型的超声定位脊柱三维空间运动分析仪———ZebrisCMS70P测量正常人以及脊柱侧凸患者脊柱形态的可靠性和重复性。方法:正常人5例,脊柱侧凸患者5例,以及5名骨科医师进行Zebris日内测量精度和日间测量精度的精度检测。已知长度和角度的物体固定放置,进行长度测量精度和角度测量精度监测以及同一受试对象不同测试者测量误差检测。结果:同一样本同一天内不同时段测量结果无显著差异,连续5天同一时段测量结果无显著差异。同一受试对象由5名骨科医师连续进行测量,结果无显著差异。长度测量精度为0.958%,角度测量精度为6.114%。结论:Zebris对脊柱形态测量可靠性和重复性较好,是一种无创性检测手段。 Objective:To report the reproducibility and reliability of Zebris CMS70P,a new ultrasonography-based system for spatial kinematics analysis of the spine and its application in measuring spine shape of scoliosis patients and healthy subjects.Method:To study reproducibility in one day,5healthy individuals and5scoliosis patients were tested five times within one day by a same orthopedic surgeon.To study reproducibility in different days,the same10subjects were tested five times at the same time in five successive days by a same orthopedic surgeon.To study precision of measuring length and degree,an object known length and degree was measured ten times successively by a same surgeon.To study difference of one subject measured by different surgeons,five surgeons measured the same object known length and degree,each surgeon measured five times.Result:No significant differences were indicated in the tests held in one day and in successive five days.Also no significant differences were indicated among the results obtained by five different surgeons from one object.The precision of measuring length was0.958%.The precision of measuring degree was6.114%.Conclusion:CMS70P is well reproducible and reliable,and is an univasive method.
出处 《中国康复医学杂志》 CAS CSCD 2003年第8期484-486,共3页 Chinese Journal of Rehabilitation Medicine
关键词 Zebris测量 脊柱侧凸 可靠性 重复性研究 Zebris scoliosis reproducibility reliability
  • 相关文献

参考文献7

  • 1Vogt L, Banzer W. Measurement of lumbar spine kinematics in incline treadmill walking[J]. Gait&Posture,1999,9: 18-23.
  • 2Vogt L, Pfeifer K, Banzer W. Comparison of angular lumbar spine and pelvis kinematics during treadmill and overground locomotion[J]. Clinical Biomechanics,2002,17: 162-165.
  • 3Markus Arand,Sylvia Neller, Lothar Kinzl, et al. The traumatic spondylolisthesis of the axis: A biomechanical in vitro evaluation of an instability model and clinical relevant constructs for stabilization[J]. Clinical Biomechanics, 2002,17:432--438.
  • 4Hiroshi Jonaia, Maria Beatriz G. Villanueva, Ayako Takataa,et al.Effects of the liquid crystal display tilt angle of a notebook computer on posture,muscle activities and somatic complaints[J]. International Journal of Industrial Ergonomics,2002,29: 219-229.
  • 5Richter M, Wippermann B. Thermann H, et al. Plantar impact causing midfoot fractures result in higher forces in Chopart's joint than in the ankle joint[J].Journal of Orthopaedic Research,2002,20:222-232.
  • 6Hermsdōrfer J, Marquardt C, Wack S, et al. Comparative analysis of diadochokinetic movements[J].Journal of Electromyography and Kinesiology,1999,9: 283-295.
  • 7Zeevi Dvir, Tamara Prushansky. Reproducibility and instrument validity of a new ultrasonography-based system for measuring cervical spine kinematics[J]. Clinical Biomechanics,2000,15:658-664.

同被引文献58

引证文献5

二级引证文献27

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部