摘要 目的 :随访耳蜗植入后出现蜗外电极或坏损电极的情况 ,评价人工耳蜗植入后电极的有效性。方法 :1998年 6月至 2 0 0 2年 8月 5 3例NucleusCI2 4M型耳蜗植入者 ,在术中及术后映射调图时进行电极阻抗测试 ,确定有无电极坏损及其坏损时间。开机时拍摄耳蜗位X线平片 ,确定电极是否完全植入。分析手术过程和术者经验对电极有效性的影响。结果 :电极出现坏损的比例为 13 2 % ,电极在术中插拔两次以上而造成电极坏损的危险度为 0 4 6 5 6。 4年累积生存率达 96 2 %。不完全植入的比例为 15 1% ,其中又以耳蜗形态异常者居多 (7/11)。结论 :NucleusCI2 4M型植入体在开机 1~ 4年内的可靠性较高。电极序列在术中插拔两次以上 ,是导致电极坏损的主要原因。患者耳蜗的纤维化。 Objective:To evaluate the efficiency of the electrodes of Nucleus CI24M through half year fellow up observation about the impaired and extra cochlear electrodes Methods: 53 patients implanted with Nucleus CI24M between June 1998 and August 2002 participated in this study The impedances of electrodes were measured intra operatively and at each mapping visits The impedance test detected the short or open circuit electrodes in the arrays The first detection time of these problem electrodes since surgery were used to calculate the cumulative survival percentage Postoperative cochlear view radiography at initial stimulation was used to determine part insertion or full insertion of the electrode array The influence of surgical procedure and surgeon's implant experiences were analyzed Results: Percentage of cases with even one impaired electrode was 13 2%; the risk ratio of re insertion for electrode impairment was 0 4656 Part implantation occurred in 15 1% patients, most of them (7/11) showed cochlea malformation The cumulative survival percentage after 4 years use was 96 2% Conclusion: Nucleus CI24M achieves a very high reliability in 1 4 years device use The re insertion during the surgery would like to cause the impairment of the electrodes The cochlea's fibrosis, malformation, and surgeon's implant experience are the main reasons of the part insertion of the electrode array
出处 《军医进修学院学报》 CAS 2003年第3期214-216,共3页 Academic Journal of Pla Postgraduate Medical School
关键词 Nucleus24型 人工耳蜗植入 电极 有效性 电阻抗 cochlear implant electrode electric impedance
  • 相关文献

参考文献7

  • 1刘伟,姜雯频,金芳华,黄以乐.结构异常耳的人工耳蜗植入术[J].听力学及言语疾病杂志,2002,10(3):198-201. 被引量:8
  • 2Von Wallenberg EL, Brinch JM. Cochlear implant reliability[J]. Ann Otol Rhinol Laryngol, 1995, 166(Suppl) : 441-443.
  • 3Lehnhardt M, von Wallenberg EL, Brinch JM. Reliability of the nucleus CI22 and CI24M cochlear implants [J]. Ann Otol Rhinol Laryngol, 2000, 185(Suppl) :14-16.
  • 4Xu J, Xu SA, Cohen LT, et al. Cochlear view: postoperative radiography for cochlear implantation[J]. Amer J Otol 2000,21:49-56.
  • 5Lambert PR, Ruth RA, Hodges AV. Meningitis and facial paresis. Implications for cochlear implantation. Arch Otolaryngol Head Neck Surg, 1987, 113:1101-1103.
  • 6Yune HY, Miyamoto RT, Yune ME. Medical imaging in cochlear implant candidates [J]. Am J Otol, 1991, 12(Suppl):11-21.
  • 7Sennarglu L, Saatci I, Aralasmak A, et at. Magnetic resonance imaging versus computed tomography in pre-operative evaluation of cochlear implant candidates with congenital hearing loss[J]. J Laryngol Otol, 2002, 116: 804-810.

二级参考文献45

  • 1Arriaga MA, Carrier D. MRI and clinical decisions in cochlear implantation[J]. Am J Otol, 1996,17:547.
  • 2Marsot DK. 3D imaging of the labyrinth:application to candidates for cochlear implant[J].Ann Radiol Paris,1992,35:44.
  • 3Lopponen H, Holma T, Sorri M, et al.Computed tomography data based rapid prototyping model of the temporal bone before cochlear implant surgery[J]. Acta Otolaryngol,1997,529:47.
  • 4Gantz BJ,McCabe BF,Tyler RS. Use of multichannel cochlear implants in obstructed and obliterated cochleas[J]. Otolaryngol Head Neck Surg,1988,98:72.
  • 5Aso S, Horiwaki S, Noda S. Surgical techniques for insertion of a multi-electrode implant into a postmeningitic ossified cochlea[J]. Am J Otol,1995,16:231.
  • 6Gibson W P. Surgical technique for inserting the cochlear multiclectrode array into ears with total neo-ossification[J].Ann Otol Rhinol Laryngol Suppl,1995,166:414.
  • 7Steenerson RL,Gary LB. Multichannel cochlear implantation in obliterated cochleas using the Gantz procedure[J]. Laryngoscope,1994,104:1071.
  • 8Lenarz T, Battmer RD,Lesinski A,et al. Nucleus double electrode array: a new approach for ossified cochleae[J]. Am J Otol,1997,18:39.
  • 9Bredberg G, Lindstrom B, Lopponen H, et al. Electrodes for ossified cochleas[J]. Am J Otol,1997,18:42.
  • 10Kiefer J, Weber A, Pfennigdorff T, et al. Scala vestibuli insertion in cochlear implantation: A valuable alternative for cases with obstructed scala tympani[J]. J Otorhinolaryngol Relat Spec, 2000,62:251.

共引文献7

同被引文献26

  • 1Lehnhardt M, von Wallenberg EL, Brinch J. Reliability of the nucleus CI22 and CI24M cochlear implants[J]. Ann Otol Rhinol Laryngol,2000, 185(Suppl): 14.
  • 2Johnson IJ, Gibbin KP, O' Donoghue GM. Surgical aspects of cochlear implantation in young children: a review of 115 cases[J]. Am J Otol,1997, 18(6 Suppl) :69.
  • 3Abbas PJ, Brown CJ, Shallop JK,et al. Summary of results using the nucleus CI24M implant to record the electrically evoked compound action potential[J]. Ear & Hearing, 1999,20:45.
  • 4Mason SM, Cope Y, Gamham J, et al. Intra - operative recordings of electrically evoked auditory nerve action potentials in young children by use of neural response telemetry with the nucleus C124M cochlear implant[J]. Br J Audiol,2001,35: 225.
  • 5Dillier N, Lai WK. Measurement of the electrically evoked compound action potential via a neural response telemetry system [ J ]. Ann Otol Rhinol Laryngol, 2002,111(5 Pt 1): 407.
  • 6Battmer R D,Laszig R, Lehnhardt E. Electrically elicited stapedius reflex in cochlear implant patients [ J ]. Ear Hear, 1990,11 (5) :370-374.
  • 7Spivak L G, Chute P M, Popp A L, et al. Programming the cochlear implant based on electrical acoustic reflex thresholes : patient performance [ J]. Laryngoscope, 1994, 104(10) :1225-1230.
  • 8Brown C J, Abbas P J, Fryanf-Bertschy H. et al. Intrao- perative and postoperative electrically evoked auditory brain stem responses in nucleus cochlear implant users: implications for the fitting process [ J ]. Ear Hear, 1994,15(2) :168-176.
  • 9Schultz A, Berger F A, Weber B P, et al. Intraoperative electrically elicited stapedius reflex threshold is related to the dosage of hypnotic drugs in general anesthesia [ J ]. Ann Otol Rhinol Laryngol, 2003,112(12) :1050-1055.
  • 10Smoorenburg G F, Willeboer C, Van Dijk J E. Speech perception in nucleus C124M cochlear implant users with processor settings based on electrically evoked compound action potential thresholds [ J ]. Audiol N eurootol, 2002,7 (6) :335-347.

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部