期刊文献+

光纤、电容液滴分析仪(二):系统设计 被引量:4

The Fiber-capacitive Drop Analyzer (Part Ⅱ):System Design
下载PDF
导出
摘要 详细介绍了光纤、电容液滴分析仪 (FCDA:Fiber- Capacitive Drop Analyzer)的系统设计方案和各组成部分的具体实现方法 ,包括液滴传感器的设计、微量供液系统的设计。 The system design of fiber capacitive drop analyzer (FCDA) and the implementation technology of its components are introduced in detail. Designs for the drop sensor, the microflow feeding pump and the signal processing circuits are presented. Computer interface for FCDA has been established.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2003年第4期352-355,359,共5页 Chinese Journal of Scientific Instrument
基金 科技部"中国新加坡联合研究计划 (0 0 3/ 1 0 1 / 0 4 ) (NSTB- MOST Joint Research Program)"资助项目 (NSTB:National Science and Technology Bureau of Singapore MOST:Ministry of Science and Technology of China.)
关键词 液滴分析仪 FCDA VFDT 液滴传感器 供液泵 信号处理 Drop analysis Drop sensor Feeding pump Signal processing
  • 相关文献

参考文献6

  • 1宋晴,张国雄,裘祖荣,陈海秀.光纤、电容液滴分析仪(一):原理与实验[J].仪器仪表学报,2003,24(4):348-351. 被引量:15
  • 2宋晴,裘祖荣,张国雄.光纤、电容液滴传感器的设计研究[J].仪表技术与传感器,2002(8):5-8. 被引量:4
  • 3宋晴,张国雄,张爱萍.调频式电容液滴传感器信号处理方式[J].电子测量技术,2003,26(2):5-6. 被引量:5
  • 4Wang C H, Augousti A T, Mason J, et al. The Capacitive Drop Tensiometer-a Novel Multianalysing Technique for Measuring the Properties of Liquids. Measurement Science and Technology, 1999, (10) : 19~24.
  • 5MeMillan N D, Finlayson O, Fortune F, et al. The Fibre Drop Analyser: a New Multianalyser Analytical Instrument with Applications in Sugar Processing and for the Analysis of Pure Liquids. Measurement Scienceand Technology, 1992, (3):746~764.
  • 6Song Qing, Zhang Aiping, Zhang Guoxiong, et al. Signal processing circuit design for the capacitive drop sensor. ISIST'2002 Proceedings, 2002,3: 652~ 658.

二级参考文献15

  • 1李华主编.MCS-51系列单片机实用接口技术[M].北京航空航天大学出版社,2002..
  • 2C H Wang et al 1999 The Capacitive Drop Tensiometer-a Novel Multianalysing Technique for Measuring the Properties of Liquids Measurement Science and Teehnology 10 19- 24.
  • 3MAXIM Corp. Max038 high- frequency waveform generator Users Manual.
  • 4INTEL Corp. Intd8254 progrmmrmble interval timer Users Manual.
  • 5Rayleigh Phil Mag, 1899,48: 321~327.
  • 6McMillan N D, Finlayson O, Fortune F, et al. The Fibre Drop Analyser: a New Multianalyser Analytical Instrument with Applications in Sugar Processing and for the Analysis of Pure Liquids. Measurement Scienceand Technology, 1992, (3): 746~764.
  • 7McMillan N D, Feeney F E, Power M J P, et al. A qualitative study of fiber drop traces on mechanically excited damped vibrations of drops of pure liquids using FFT analysis. Instrumentation Science and Technology, 1994,22(4): 375~395.
  • 8McMillan N D, Davern P, Lawlor V, et al. The instrumental engineering of a polymer fibre drop analyzer for both quantitative and qualitative analysis with special reference to fingerprinting liquids. Colloids and Surfaces A,1996,114.75~97.
  • 9Wang C H, Augousti A T, Mason J, et al. The Capacitive Drop Tensiometer-a Novel Multianalysing Technique for Measuring the Properties of Liquids. Measurement Science and Technology, 1999, (10)19~24.
  • 10Tait T Phil Mag, 1864,27:176.

共引文献17