Kappa 700型双腔起搏器的临床随访

Clinical Oberservation of Dual Chamber Pacemaker With Ventricular Autocapture Function
下载PDF
导出
摘要 观察具有自动夺获功能的双腔起搏器 (Kappa 70 0 )置入后参数的变化和安全性情况。随访 1 3例置入Kappa70 0型起搏器患者 ,观察术中、术后 1周及术后 1 ,3,6个月心室起搏阈值、输出电压、输出脉宽、电极阻抗、R波振幅的变化 ,了解起搏器的工作情况。术后测得的起搏阈值较术中明显升高 ( 0 .71± 0 .2 3Vvs 0 .39± 0 .0 6V ,P <0 .0 5 ) ,术后不同时间测得的起搏阈值无明显差异。R波振幅术中、术后无明显差异。术后阻抗较术中明显降低 ( 62 5 .7± 1 2 3.0Ωvs 894.3± 1 90 .3Ω ,P <0 .0 5 ) ,术后 1个月后的阻抗基本稳定。起搏器自动夺获功能打开后 ,平均输出电压为 0 .96~ 1 .1 6V ,平均输出脉宽 0 .32~ 0 .34ms,平均心房感知灵敏度 0 .71~ 0 .83mV ,心室感知灵敏度 3.82~3.91mV。随访期间起搏、感知功能正常 ,无误感知现象。具有自动夺获功能的双腔起搏器输出电压低 ,安全可靠。 To observe the changing parameters and the safety of dual chamber pacemaker with ventricular autocapture function, ventricular stimulation threshold, output amplitude, impedance, R wave amplitude, and the status of pacemaker were measured in the thirteen patients implanted Medtronic Kappa 700 at the time of implantation and 1st week, 1st months, 3rd months, 6th months after implantation. Six patients had ambulatory electrocardiographic records in the three months after implantation. The ventricular pacing threshold increased significantly after implantation(0.71±0.23 V vs 0.39 ±0.06 V , P <0.05), and did not any change during follow up. R wave had no change. The impedence decreased significently after implantation(625.7±123.0 Ω vs 894.3±190.3 Ω, P <0.05), there were no difference after 1st months. The mean output amplitude was 0.96-1.16 V, output amplitude 0.32-0.34 ms, atrial sensitivity 0.71-0.83 mV, ventricular sensitivity 3.82 -3.91 mV. All pacemaker worked effectively and safely during follow up for 7-31months. The result suggest that the dual chamber pacemaker with ventricular autocapture function could safely work with low output amplitude.
出处 《中国心脏起搏与心电生理杂志》 2003年第4期267-269,共3页 Chinese Journal of Cardiac Pacing and Electrophysiology
关键词 Kappa700型双腔起搏器 临床随访 心律失常 心血管病学 自动夺获功能 Cardiology Arrhythmias Dual chamber pacing Autocapture
  • 相关文献

参考文献7

  • 1杨杰孚.Kappa 700型起搏器的阈值管理及感知保障功能[J].中国心脏起搏与心电生理杂志,2002,16(1):5-7. 被引量:8
  • 2郭继鸿.起搏器的自动化功能——兼谈Kappa700起搏器(续)[J].中国心脏起搏与心电生理杂志,2002,16(6):417-423. 被引量:18
  • 3Simeon L, Duru F, Fluri M, et al. The impact of automatic threshold tracking on pulse generation longevity in patients with different chronic stimulation thresholds[J]. PACE, 2000,23 : 1788.
  • 4Dohrmann MI, Goldschlager NF. Myocardial stimulation threshold in patients with cardiac pacemakers: Effect of physiologic variable,pharmacologic agents and lead electrodes [ J ]. Cardiol Clin, 1985,3:527.
  • 5Feld GK, Love C J, Camerlo J, et al. A new pacemaker algorithm for continuous capture verification and automatic threshold determination[J]. PACE, 1992, 15:171.
  • 6Madrid AH, Olague J, Cercas A, et al. A prospective multicenter study on the safety of a pacemaker with automatic energy control: Influence of the electrical factor on chronic stimulation threshold [ J ].PACE, 2000, 23:1359.
  • 7Clarke M, Liu B, Schuller L, et al. Automatic adjustment of pacemaker stimulation output correlation with continuously monitored capture thresholds: A multicenter study[J]. PACE, 1998,21:1567.

共引文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部