期刊文献+

激光探针等离子体质谱测定NIST硅酸盐玻璃标样中痕量元素

Determination of Trace Elements of NIST Silicate Glass by Laser Probe Inductively Coupled Plasma Mass Spectrometry
下载PDF
导出
摘要 采用激光探针等离子体质谱技术 ,以 42 Ca为内标 ,经预熔样处理 ,获得 NIST硅酸盐玻璃标样中 2 4种痕量元素 (Rb、Sr、REE、Th、U等 )的含量。结果表明 :方法的检出限在取样孔径 5 0 μm条件下可达 <1 0 -7(U、Th)~ <1 0 -6(REE) ,以 NIST61 2的一点为标准测定其它点 ,大多数元素的精密度优于± 1 0 % ;以NIST61 2为标准测定 NIST61 0 ,大多数元素的准确度优于± 1 0 %。 The concentrations of the twenty-four species of trace elements such as Rb, Sr, REE, Th and U of NIST silicate glass are obtained with laser probe inductively coupled plasma mass spectrometry technique. The signal variation of representative isotopes such as 42 Ca, 139 La, 175 Lu, 208 Pb and 238 U versus ablation time for a typical sampling spot on NIST612 is firstly investigated. It is shown that the signal of measured isotopes is stable after 30 seconds and the inter-element fractionation between Ca and other elements is negligible after 20 seconds. 42 Ca was chosen as an internal standard. Detection limits determined by 3σ of carrier gas blank are found to be lower than 10 -7 for U and Th to lower than 10 -6 for REE with an ablation crater of about 50 μm. The precision is better than 10% for most elements such as REE, U, Th and Pb according to analyses of 6 sampling spot of NIST612 using NIST612 itself as a calibrated standard. The accuracy is also better than 10% for most elements according to analyses of NIST610 using NIST612 as a calibrated standard.
作者 储著银
出处 《质谱学报》 EI CAS CSCD 2003年第3期389-393,共5页 Journal of Chinese Mass Spectrometry Society
关键词 激光探针等离子体质谱 测定 NIST硅酸盐玻璃 痕量元素 质谱学 微区分析 样品 mass spectrometry determination of trace elements laser probe inductively couple plasma mass spectrometry NIST silicate glass microanalysis
  • 相关文献

参考文献9

  • 1储著银,孙敏,周新华.激光进样等离子体质谱技术及其在地球科学中的应用[J].岩矿测试,1998,17(2):152-158. 被引量:6
  • 2Norman MD, Pearson NJ, Sharma A, et al.Quantitative Analysis of Trace Elements in Geological Materials by Laser Ablation ICPMS:Instrumental Operating Conditions and Calibration Values of NIST Glasses[J]. Geostandards Newsletter, 1996, 20(2) :247.
  • 3Pearce NJG, Perkins WT, Abell I, et al. Mineral Microanalysis by Laser Ablation Inductively Coupled Plasma Mass Spectrometry[J]. J Anal AtSpectrom, 1992, 7:53-57.
  • 4Jackson KE, Longerich HP, Dunning GR, et al.The Application of Laser Ablation Microprobe Inductively Coupled Plasma Mass Spectrometry(LAM-ICP-MS) to in Rare Earth Trace Element Determination in Minerals[J]. Canadian Mineralog, 1992, 30:1049-1064.
  • 5Jeffries TE, Perkins WT, Pearce NJG. Comparison of Infrared and Ultraviolet Laser Probe Microanalysis Inductively Coupled Plasma Mass Spectrometry in Mineral Analysis[J]. Analyst,1995, 120(5):1365-1371.
  • 6Feng R, Machado N, Ludden J. Lead Geochronology of Zircon by Laser Probe-Inductively Coupled Plasma Mass Spectrometry (LP-ICPMS)[J]. Geochimica et Cosmochimica Acta, 1993,57:3479-3486.
  • 7Fryer BJ, Jackson SE, Longerich HP. The Application of Laser Ablation Microprobe-Inductively Coupled Plasma-Mass Spectrometry (LAM-ICPMS) to in Situ(U)-Pb Geochronology[J]. Chem Geol, 1993, 109:1-8.
  • 8Hirata T, Nesbitt RW. U-Pb Isotope Geochronology of Zircon:Evaluation of the Laser Probe-Inductively Coupled Plasma Mass Spectrometry Technique[J]. Geochimica et Cosmochimica Acta, 1995, 59(12):2491-2500.
  • 9Fryer B J, Jackson SE, Longerich HP. The Design, Operation and Role of the Laser-ablation Microprobe Coupled With an Inductively Coupled Plasma-Mass Spectrometer (LAM-ICP-MS) in the Earth Sciences [J]. Canadian Mineralog,1995, 33:303-312.

二级参考文献6

共引文献5