液氮保存同种带支架人工瓣膜流体力学测试 被引量:4

In vitro hemodynamic evaluation of the cryopreserved stented homograft bioprosthetic heart valve
原文传递
导出
摘要 目的 利用脉动流模拟实验装置测试液氮保存同种带支架瓣膜流体力学性能 ,同时与国产perfect牛心包人工瓣膜对比研究。 方法 采集同种瓣膜 ,缝制成 2 1# 、2 3# 、2 5 # 同种带支架主动脉和同种带支架肺动脉人工瓣膜 ,经液氮保存 ,使用国产脉动流实验装置测试瓣膜流体力学性能 ,采用ISO/FDA评估标准 ,分别测量各流量下的跨瓣压差、有效瓣口面积 (EOA)和回流比。并与相应型号国产 perfect牛心包人工瓣膜对比研究。结果  2 1# 、2 3# 、2 5 # 同种带支架主动脉和肺动脉瓣膜的跨瓣压差和回流百分比差异无显著性 ,但较 perfect瓣大。同种带支架主、肺动脉瓣膜的EOA差异无显著性 ,但较 perfect牛心包生物瓣膜的略小些。同种带支架瓣膜实际开口面积 (AOA)小于 perfect牛心包瓣 ,但有效瓣口面积 /实际开口面积比值无差别。结论 同种带支架主动脉和肺动脉人工瓣膜流体力学性能满意 ,同种带支架主动脉瓣膜与同种带支架肺动脉瓣膜流体力学性能无差别。 Objective: To evaluate hemodynamic function of the cryopreserved stented homograft bioprosthetic valve , compared to Perfect bovine pericardial bioprosthetic heart valve. Methods: Three sorts of stented homograft aortic and pulmonary valve (21 #, 23 #, 25 #) preserved by liquid nitrogen were prepared. In vitro pulsatile flow tests were performed on the valve in accordance with ISO/FDA guideline. Effective orifice area (EOA), transvalve pressure gradient and regurgitation ratio were recorded at various flow volume, compared with Perfect bioprosthetic valve. Results: There was no difference between the transvalve pressure gradient of the stented homograft aortic and pulmonary valve, but they were higher than Perfect bioprosthetic valve. The stented homograft aortic and pulmonary valve showed the larger regurgitation ratio than perfect valve, but there was no difference between two types of valve. EOA of the stented homograft aortic and pulmonary valve were smaller than Perfect valve's, there was no discrepancy between them. Though actual orifice area (AOA) of the stented homograft valve was smaller than Perfect's, but EOA/AOA ratio was no difference between two types of valve.Conclusion: The stented homograft valve has satisfactory hemodynamic performance. There is no significant discrepancy between homograft aortic and pulmonary valve's hemodynamic function.
出处 《中华胸心血管外科杂志》 CSCD 北大核心 2003年第4期232-234,共3页 Chinese Journal of Thoracic and Cardiovascular Surgery
基金 国家高技术研究发展计划 (863计划 )基金资助(2 0 0 1AA2 1 60 61 )
关键词 液氮保存 人工瓣膜 流体力学 测试 低温保藏 Heart valve prosthesis Cryopreservation Hemodynamic
  • 相关文献

参考文献9

  • 1Kiddin JK, Smith D, Novick W, et al. Long-term function d crycpreserved aortic homografts. A ten-year study. J Thorac Catdiovasc Surg,1993,106:154 - 166.
  • 2Jin XY, Zhang ZM, Gibson DG, et 81. Effects d valve substitute on changes in left ventricular function and hypertrophy after aortic valve replacement. Ann Thorac Stag, 1996,62:683 -690.
  • 3Yacoub M, Rasmi NR, Sundt TM, et al. Fourteen-year experience with homovital homografts for aortic valve replacement. J Thorac Cardiovasc Surg, 1995,110:186- 194.
  • 4Reul H, Talukder N, Muller EW. Fluid mechanies of the natural mitral valve. J Biomech, 1981,14:361 - 372.
  • 5Fisher J. Hydrodynamic function d ten prosthetic heart valves in the aortic position. CAin Phys Physiol Mees, 1988, 19:307-317.
  • 6Walker DK, Scotten LN, Brownlee RT. The Mitrflow pericarrdial heart valves. In vitro assessment over a range of sizes in aortic and mitral positions. Scan J Thora Cardiovasc Surg, 1985,19:131 - 138.
  • 7Wright JT. An in-vitro assestment of the hydraulic characteristics of the mark II Abrams-Luca mitral valve prosthesis.Thorax,1977,32:296-302.
  • 8Heiliger R, Geks J, Richter H, et al. Regurgitation and orifiee area of bioprostheses for miral valve replacement. Biomed Tech(Bed), 1985,30:182- 186.
  • 9Knott E, Reul H, Knoch M, et al. In vitro comparison ofaortic heart valve prostheses. Part I. Mechanical valves. J Thorac Cardiovase Surg, 1988,96:952 - 961.

同被引文献64

引证文献4

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部