期刊文献+

流动注射在线萃取色谱分离原子吸收光谱法测定痕量铂 被引量:13

Flow-injection on-line extraction chromatography for determination of trace platinum by flame atomic absorption spectrometry
下载PDF
导出
摘要 研究了流动注射在线分离富集原子吸收光谱法测定痕量铂的方法。以自制的GDX501 TBP萃取树脂为微型分离柱,在优化后的分离富集条件下,进样时间为60s,洗脱时间为45s。在线分离测定时间为3min,方法检出限为0.25μg/L,线性范围10~600μg/L,加标回收率为97.8%~103 6%,相对标准偏差3.49%~4.25%。方法已用于矿物管理样中铂的测定。 A system of flow injection online extraction chromatographic separation and determination of trace platinum by flame atomic absorption spectrometry was studied. The microcolumn was packed with GDX501TBP extraction resin as separation column. In the 60 s sampling and the 45 s of elution, platinum ions were absorbed and eluted. The online separation time was only 3 min. The calibration curve was linear for Pt in the range of 10~600 μg/L with a detection limit of 0.25 μg/L, recoveries of 97.8%~1036%, and precision of 3.49%~4.25% RSD. It was applied to the determination of platinum in mineral samples.
出处 《分析试验室》 CAS CSCD 北大核心 2003年第4期8-10,共3页 Chinese Journal of Analysis Laboratory
基金 湖南省教育厅自然科学基金资助(02C576)
关键词 流动注射 在线分离 萃取色谱 分离 原子吸收光谱法 测定 痕量元素 Flow injection Extraction Chromatography AAS Platinum