摘要
建立了分离检测环境空气中酚类化合物的超高效液相色谱法。采用ACQUITY UPLCHSS T3色谱柱进行分离,柱温30℃,以超纯水(含0.5%乙酸)和乙腈(含0.5%乙酸)作为流动相进行梯度洗脱,流速0.2 m L/min,进样量5.0μL,280 nm波长条件下紫外检测;可以实现环境空气中11种酚类化合物的分离测定。结果表明,酚类含量在0.5~40μg/m L范围内线性关系良好,相关系数均大于0.999;在采样体积为25 L时,方法检出限范围为4~39μg/m^3,相对标准偏差0.39%~5.0%,质控样测定相对误差在1.5%~3.8%之间。本方法分离测定环境大气中酚类时准确性好、灵敏度高、操作简便。
A new method was developed to determinate 11 phenolic compounds in the air by ultra-high performance liquid chromatography equipped with an ACQUITY UPLCHSS T3 column and an ultraviolet detector. The column temperature was maintained at 30 ℃. The mobile phases were ultrapure water(containing 0. 5% acetic acid) and acetonitrile(containing 0. 5% acetic acid). The flow rate of mobile phase was 0. 2 m L/min. The injection volume was 5. 0 μL. The linear range of 11 phenolic compounds were from 0. 5 to 40 μg/m L and the correlation coefficients were all higher than 0. 999. The method detection limits ranged from 4 to 39 μg/m^3 when the sampling volume was 25 L. The relative standard deviations were in the range of 0. 39% to 5. 0%. The relative errors of determination results for quality control sample ranged from 1. 5% to 3. 8%. The results showed that the method had the advantages of good accuracy,high sensitivity and easy operation.
出处
《应用化工》
CAS
CSCD
北大核心
2017年第3期589-591,596,共4页
Applied Chemical Industry
基金
国家自然科学基金(41302206)
中央高校基本科研业务费专项资金项目(310829161001)
陕西省矿产资源勘查与综合利用重点实验室基金(2014HB008)
长安大学中央高校基金陕西省环境保护水土污染与修复重点实验室开放基金(310829161130)
关键词
超高效液相色谱法
环境空气
酚类
ultra-high performance liquid chromatography
environmental air
phenolic compounds