摘要 针对加速器不适宜在实验室使用且发出的射线会伤害人体等问题,为方便考察仪器性能,采用标准光源模拟加速器照射的方法,对矩阵剂量仪半导体探测器进行性能测试。经过刻度之后,矩阵剂量仪能够完成在加速器照射下的实际测量。实验结果表明:采用标准光源模拟加速器照射矩阵剂量仪,对半导体探测器进行性能测试的技术是可行的。 Accelerator is not suitable for use in the laboratory and its emitting rays will harm the human body. In order to test the instrument performance, semiconductor detectors of matrix dosimeter are tested using standard light source to simulate accelerator. Matrix dosimeter can complete the actual measurement in the accelerator radiation after calibration. As we can conclude,performance testing technology of semiconductor detectors is feasible,that standard light source is used to simulate accelerator.
出处 《中国测试》 CAS 北大核心 2015年第3期117-120,共4页 China Measurement & Test
基金 国家重大科学仪器设备开发专项(2013YQ090811)
关键词 加速器 矩阵剂量仪 半导体探测器 标准光源 accelerator matrix dosimeter semiconductor detector standard light source
  • 相关文献

参考文献4

  • 1方方等编著,贾文懿主编.核地球物理仪器[M]. 原子能出版社, 1998
  • 2复旦大学等编,吴治华等.原子核物理实验方法[M]原子能出版社,1997.
  • 3C. Bozzi,V. Carassiti,A.Cotta Ramusino,S. Dittongo,M. Folegani,L. Piemontese,B.K. Abbott,A.B. Breon,A.R. Clark,S. Dow,Q. Fan,F. Goozen,C. Hernikl,A. Karcher,L.T. Kerth,I. Kipnis,S. Kluth,G. Lynch,M. Levi,P. Luft,L. Luo,M. Nyman,M. Pedrali-Noy,N.A. Roe,G. Zizka,D. Roberts,D. Barni,E. Brenna,I. Defendi,A. Forti,D. Giugni,F. Lanni,F. Palombo,V. Vaniev,A. Leona,E. Mandelli,P.F. Manfredi,A. Perazzo,V. Re,C. Angelini,G. Batignani,S. Bettarini,M. Bondioli,F. Bosi,G. Calderini,M. Carpinelli,F. Dutra,F. Forti,D. Gagliardi,M.A. Giorgi,A. Lusiani,P. Mammini,M. Morganti,F. Morsani,E. Paoloni,A. Profeti,M. Rama,G. Rampino,G. Rizzo,F. Sandrelli,G. Simi,G. Triggiani,S. Tritto,R. Vitale,P. Burchat,C. Cheng,D. Kirkby,T. Meyer,C. Roat,M. Bona,F. Bianchi,F. Daudo,B.Di Girolamo,D. Gamba,G. Giraudo,P. Grosso,A. Romero,A. Smol,P. Trapani,D. Zanin,L. Bosisio,G.Della Ricca,L. Lanceri,A. Pompili,P. Poropat,M. Prest,C. Rastelli,E. Vallazza,G. Vuagnin,C. Hast,E.P. Potter,V. Sharma,S. Burke,D. Callahan,C. Campagnari,B. Dahmes,A. Eppich,D. Hale,K. Hall,P. Hart,N. Kuznetsova,S. Kyre,S. Levy,O. Long,J. May,J. Richman,W. Verkerke,M. Witherell,J. Beringer,A.M. Eisner,A. Frey,A. Grillo,M. Grothe,R. Johnson,W. Kroeger,W. Lockman,T. Pulliam,W. Rowe,R. Schmitz,A. Seiden,E. Spencer,M. Turri,M. Wilder,E. Charles,P. Elmer,J. Nielsen,W. Orejudos,I. Scott,J. Walsh,H. Zobernig.The design and construction of the BaBar silicon vertex tracker[J]. Nuclear Inst. and Methods in Physics Research, A . 2000 (1)
  • 4Kawasaki,T.The Belle silicon vertex detector. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment . 2002

共引文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部