摘要
针对含硼富燃料推进剂金属含量大,燃烧后残渣导电,用普通燃速仪测试不易获得数据的问题,对靶线法测速仪的计时系统进行了改进。改进后的计时系统对测试环境的要求降低,从而提高了测试效率和精度。
Boron-based fuel-rich propellent contains metal and its combustion residue is of electro-conductive, so the burning rate data of the propellant are difficultly gained by use of an ordinary strand burner method. The problem is solved by improving the recording time system of the testing apparatus. The results show that the new timing device is less dependent on the testing environment conditions, so its testing efficiency and reliability are improved.
出处
《固体火箭技术》
EI
CAS
CSCD
北大核心
2003年第3期74-75,共2页
Journal of Solid Rocket Technology
基金
国防科技重点实验室基金51470020101HT4201
关键词
含硼富燃料推进剂
燃速仪
测试精度
计时系统
boron-based fuel-rich propellant
burning rate
strand burner
test set
timing device