摘要
采用粉末压片-X射线荧光法,测量了水系沉积物、土壤、岩石等国家一级标准物质。对用LiF(200)晶体在20.70°~48.00°衍射角(2θ)得到的X射线背景数据进行了研究。以30.97°和25.70°(2θ)为参考背景角度,进行了背景曲线的幂函数拟合,背景相对强度的拟合回代值与实测值的相对误差小于2.6%,多数小于1%。利用该拟合函数,可以计算出20.70°~48.00°内任意两个角度下的背景比值(背景系数)。因此,实验中可以根据情况选取合适的公共背景角度,并用拟合函数计算各元素谱峰角度处的背景系数。可采用公共背景法的元素数为13个左右,适当延长公共背景点的测量时间,可以降低背景的统计涨落。
32 national geological reference materials namely GBW 07301~07312,GBW 07401~07408,GBW 07103~07114, were measured by Xray fluorescence spectrometry with pressed pellet sample preparation technique. The background data for 2θ angles ranged from 20.70° to 48.00° collected by using LiF (200) crystal were studied. Power function fitting of background was carried out with background intensities measured at angles 30.97° and 25.70°(2θ) as reference background. The relative deviations of the calculated background ratios and the measured ones were within 2.6% while most of them less than 1%.The fitted power function can be used to calculate background ratios(coefficients) of any two angles in the 2θ angle range of 20.70° to 48.00°. Thus common background method can be used conveniently. In this 2θ angle range, about 13 elements can be measured by common background method with high efficiency.
出处
《岩矿测试》
CAS
CSCD
北大核心
2003年第3期161-164,共4页
Rock and Mineral Analysis
基金
国土资源部地质大调查项目(DKD9904017)
关键词
地质样品
X射线
荧光分析
背景相关曲线
X-ray fluorescence spectrometry
geological materials
background curve