期刊文献+

基于射频感应控制的掌指人工关节研究 被引量:4

STUDY ON THE ARTIFICIAL MP JOINT UNDER THE CONTROL OF RF FIELD
下载PDF
导出
摘要 针对目前临床上存在大量掌指臂丛神经损伤患者无法医治的状况 ,提出了通过植入带有驱动电机的掌指人工关节来行使掌指功能的治疗方案。文中研究了植入式人工关节的结构、机械传动、控制信号的获取和传输等。为解决驱动电机的电源问题 ,作者提出了运用电磁感应的经皮能量传输技术来驱动植入电机的旋转 ,利用反映掌指动作意识的肌电信号来控制射频磁场的发射 ,使得体内线圈感应体外射频磁场后驱动电机转动。最后研究了一套临床关节植入的手术方案 ,并在三例尸体上进行了功能实验 ,验证了植入的人工关节能满足人体掌指的基本功能。 In order to solve the problem that presently a number of patients with brachial plexus disease resulting in injury of metacar pofhalangeal joint are cureless, the paper describes a scheme that performs the functions of MP joint by implanting artificial joint with electrical drive. In this paper, some issues such as the structure of the implanting artificial joints, mechanical transmission and the acquest and transfer of controlling signal are discussed. To solve the problem of power supply for the electrical drive, the author adopts the transcutaneous energy transmission technology that makes use of electromagnetic induction to impel the implanted electrical drive to rotate. By detecting the EMG that reflects the behavioral consciousness of MP joint, the external RF field was irradiated and inducted by the implanted coils to control the rotation of the electrical drive. An operating scheme of clinical joint implantation has been researched and experimented on three cadavers. The results revealed that the implanted artificial joints could fulfill the daily function of humans MP joint.
出处 《中国生物医学工程学报》 EI CAS CSCD 北大核心 2003年第5期418-427,共10页 Chinese Journal of Biomedical Engineering
基金 国家 8 63计划的资助 ( 0 5 15 )
关键词 经皮能量传输 掌指关节 植入式 Electromagnetism Implants (surgical) Joint prostheses
  • 相关文献

参考文献6

  • 1王保华,袁冲明.体内植入电子系统(二)[J].中国医疗器械杂志,1989,13(6):352-357. 被引量:2
  • 2王保华.植入式电子系统中射频线圈的失配分析[J].上海生物医学工程,1996,17(1):11-15. 被引量:5
  • 3Sokal NO, Sokal A. Class E-A new class of high-efficiency tuned single-ended switching power amplifiers[J ]. IEEE Journal of Solid-State Circuits, 1975,10(3) : 168-176.
  • 4Troyl PR, Scgwan MA. Class E driver for transcutanecous power and data link for implanted electronic device[J]. Medical & Biological Engineering & computing, 1992,30( 1 ) : 69-75.
  • 5Ghahary A, Cho BH. Design of transcutaneous energy transmission system using a series resonant converter[J ]. IEEE Transactions on Power Electronics, 1992,7(4) :261-279.
  • 6Zierhofer CM, Hochmair E. High-efficiency coupling-insensitive transeutaneous power and data transmission via an inductive link[J]. IEEE Transactions on Biomedical engineering, 1990,37(7) :716-722.

二级参考文献1

共引文献5

同被引文献37

引证文献4

二级引证文献30

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部