期刊文献+

超声相控阵系统中相控发射与同步的实现 被引量:14

REALIZATION OF PHASE-CONTROLLED TRANSMISSION AND SYNCHRONIZATION OF ULTRASONIC PHASED ARRAY SYSTEM
下载PDF
导出
摘要 研制的超声相控阵实验系统采用数字方式控制各阵元的超声发射相位 ,能够得到很高的精度和稳定性。阐明该系统各阵元间的发射同步这一重要环节的实现方法并进行详细误差分析。结果表明 ,该方法对于规模不太复杂的相控阵系统具有足够的精度 。 The experimental ultrasonic phased array system developed was described, in which the ultrasound transmission phase of each element was controlled digitally, and high accuracy and stability were obtained. The realization of the significant transmission synchronization of elements was described and error analysis was given in detail. Result showed that the synchronization method adopted was accurate enough for moderate scale ultrasonic phased array system. A primary scheme was proposed for the synchronization of larger-scale one.
机构地区 清华大学机械系
出处 《无损检测》 2003年第10期507-510,共4页 Nondestructive Testing
基金 清华大学 985无损检测与质量控制基金资助项目
关键词 超声检验 相位控制 同步 误差分析 超声相控阵 实验系统 超声发射 Ultrasonic inspection Phase control Synchronization Error analysis