摘要
对常规傅里叶变换发光测量方法进行改进 ,引入双调制技术 ,在迈克耳孙干涉仪一级调制的基础上 ,引入更高频率调制锁相测量手段。通过这一测量方法的引入 ,并通过对调制频率和带通滤波器等测量参数的优化 ,基本消除了室温背景的黑体辐射在 4μm~ 5 μm以上区域对光致发光测量带来的严重干扰 ,在 10 μm长波红外波段得到了无室温背景黑体辐射影响的光致发光光谱 ,从而将光致发光测量推至 5
The high frequency double modulation and lock-in technique were attached to the normal Fourier transform infrared spectrometer to eliminate the huge influences of the room temperature background blackbody emission on the measurement of the infrared photoluminescence spectra. The experimental method was optimized by adjusting the modulation frequency, choosing the suitable band-pass filter and phase correction. The photoluminescence spectra in 10 μm long wave infrared band without the influences of the background blackbody emission were obtained.
出处
《光学学报》
EI
CAS
CSCD
北大核心
2001年第12期1489-1492,共4页
Acta Optica Sinica
基金
国家自然科学基金 (6 187 6 0 42
6 982 30 0 2 )