摘要 文中利用透射实验测量了全血的近红外光谱,用偏最小二乘法(PartialLeastSquare,简写为PLS)选择包含有葡萄糖特征峰的谱段,建立了回归模型,得到了葡萄糖浓度的拟合曲线,分析拟合曲线与真值的关系,相关系数达到0.9999,误差小于0.13mmol/l。实验结果表明:用近红外光谱的方法可以实现全血血糖浓度的测量。 Nearinfrared spectrum of whole blood is measured by nearinfrared spectrum transmission experiment. Consequently the calibrating model is set up within the spectrum band where the characteristic peak of glucose is found by the Partial Least Square (PLS) method. And then the curve of glucose concentration is obtained. According to the relation between the simulating curve and the real value, it is achieved correlation coefficient up to 0.9999 and error less than 0.13mmol/l. The result shows that nearinfrared technology is effective way to measure the concentration of blood glucose.
出处 《激光与红外》 CAS CSCD 北大核心 2003年第5期328-330,共3页 Laser & Infrared
基金 中科院长春光机与物理研究所二期创新项目
关键词 全血 血糖浓度 近红外光谱 最小二乘法 糖尿病 whole blood glucose concentration near-infrared spectrum PLS
  • 相关文献

参考文献7

  • 1张玲.PLS定标法在近红外光谱分析仪中的应用研究[J].光学精密工程,2000,8(3):238-241. 被引量:18
  • 2沈耀春,王春,H.A.MacKenzie,陆祖宏.一种新型的检测血糖含量的方法[J].南京大学学报(自然科学版),2000,36(5):632-635. 被引量:13
  • 3Stephen F Malin,Timothy L Ruchti, Thomas B Blank, et al. Noninvasive Prediction of Glucose by Near-infrared Diffuse Reflection Spectroscopy [ J ]. Clinical chemistry, 1999,45 (9) : 1651 - 1658.
  • 4Jeffrey W Hall, Alan Pollard. Near-Infrared Spectrophotometry: A New Dimension in Clinical Chemistry [ J ]. Clinical Chemistry, 1992, 38 (9) : 1623 - 1631.
  • 5Michael J Mcshane, Gerard L, Clifford Spiegelman. Variable selection for quantitative determination of glucose with near-infrared spectroscopy[ A]. SPIE, 1997,2952 :159 - 197.
  • 6Vidi A Saptari, Kamal Youcef-toumi. Sensitivity analysis of near infrared glucose absorption signals : toward noninvasive blood glucose sensing [ A ]. SPIE, 2000,4163 : 45-54.
  • 7H Michael Heise, Andreas Bittner. Blood glucose assays based on infrared spectroscopy: Alternatives for medical diagnostics [ A ]. SPIE, 1998, 3257:2 - 11.

二级参考文献7

  • 1[1]Norris K H, Williams P C. Optimization of mathematical treatments of raw near-infrared signal in the measurement of protein in hard red spring wheat. Ⅰ. influence of particle size[J]. Cereal Chemistry, 1984, 61(2): 158-165.
  • 2[2]Geladi Paul, Kowalski Bruce R. Partial least-squares regression: a tutorial[J]. Analytica Chimica Acta,1986, 185: 1-17.
  • 3[3]Geladi Paul, Kowalski Bruce R. An example of 2-block predictive partial least-squares regression with simulated data[J]. Analytica Chimica Acta, 1986, 185: 19-32.
  • 4[4]Haaland David M. Thomas Edward V. Partial least-squares methods for spectral analyses. 1. relation to other quantitative calibration methods and the extraction of qualitative information[J]. Anal. Chem., 1988, 60: 1193-1202.
  • 5[5]Mcshane Michael J, Cote Gerard L, Spiegelman Clifford H. Assessment of partial least-squares calibration and wavelength selection for complex near-infrared spectra[J]. Applied Spectroscopy, 1998, 52(6): 878-884.
  • 6赵学萍,程玲.广东省公共图书馆多媒体信息网的建设与发展[J].图书馆论坛,1999,19(4):67-69. 被引量:3
  • 7胡玉文.论新环境下中国图书馆员的继续教育[J].中国图书馆学报,1999,25(4):91-93. 被引量:67