摘要 TechniconRA - 1 0 0 0全自动生化分析仪上惰性液的回收方法简单易行 ,经济实用。用回收的惰性液和原机进口的惰性液进行了多种对比实验 ,结果令人满意。其中用测定总胆固醇 (TC)、丙氨酸氨基转移酶 (ALT)和尿素氮 (UN)的t值分别为 1 52 6 ,1 930 ,1 590 ,P均大于 0 0 5 ,说明用回收的惰性液在Techni conRA - 1 0 0 0全自动生化分析仪上应用是可行的。 The method of Retrieving the Randon Access Fluid is simple,practical and economical A lot of comparing experiments have been done between original Randon Access Fluid and retrieved Randon Access Fluid on Technicon RA-1000 biochemical autoanalyzer The results are ideal for example,TC,t=1 526,P>0 05 ALT,t=1 930,P>0 05 UN, t=1 590, P>0 05 So it proves that retrieved Randon Access Fluid is available on Technicon RA-1000 biochemical autoanalyzer
作者 贺先奇
机构地区 解放军
出处 《医疗装备》 2003年第9期7-9,共3页 Medical Equipment
关键词 TeehnieonRA—1000型 全自动生化分析仪 惰性液 交叉污染 Randon Access Fluid on Technicon RA-1000 method of Retrieving biochemical autoanalyzer